Press ENT key to store it.
->
->
(In case it is not good,
will come out.)
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
(Currently the value is indicated.)
ex) When DAC CAPA value sets at 500.0, analog output voltage will be
10V
.
6-10. Setting the number of buffer
As setting Display speed and the number of A/D Sampling , it can set up 1~32.
According to 6-1, Press Function key and then Up key to be
.
->
Press ENT key to choose the number of buffer.
Press Up key to change the required number.
…….
Fast --------------------------- Slow
Press ENT key to store it.
->
->
(In case it is not good,
will come out.)
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
(Currently the value is indicated.)
6-11. Setting Auto Tracking Range
As setting ZERO Operation Range (Auto zero), it set up 0, 1, 2, 5, 10, 20, 50, 100.
※ This function must use 6-12 Time Mode at the same time.
According to 6-1, Press Function key and then Up key to be
.
->
Press ENT key to choose Auto zero.
Press Up key to change the required number.
…….
Press ENT key to store it.
12
->
->
(In case it is not good,
will come out.)