37
Upon coming across any problem in operation, follow the instructions below to check and
repair the problem, and/or get in touch with your representative.
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS
Problem
Probable cause
Solution
Delivery Unit
-Handpiece is not working.
- Compressor disconnected.
- Plug the compressor in.
-Handpiece with low speed.
-Inlet pressure below specified
(80 PSI).
-Adjust inlet pressure (80
PSI).
-No water from handpiece
spray.
-Insufficient air pressure
from compressor.
-Reservoir run out of water.
-Closed terminal.
- Adjust air flow.
- Put filtered water in reservoir.
- Open terminal.
- No water from syringe.
-Reservoir run out of water.
-Compressor disconnected.
-Put filtered water in reservoir.
-Plug compressor in.
- When Flush System is
operated no disinfectant come
from handpiece terminals.
-Flush System reservoir run out
of water.
- Chair fuse burned.
-Main or chair switch is off.
-Put disinfectant in the
reservoir.
- Turn off the chair from mains
power and request a Technician
presence.
-Switch main/chair switch on.
- X ray view does not work.
-Chair’s fuse burned.
-Main switch is off.
Turn off the chair from mains
power and request a Technician
presence.
-Switch the main switch on.
Curing Light
- Equipment’s not working.
-Power cut.
-Chair’s fuse burned.
-Check power supply.
-Turn off the chair from mains
power and request a Technician
presence.
- Equipment is not polymerizing
resins.
-Resin is not appropriate for
LED’s photopolymerizer wave
length range.
-Resin residue in the light
conductor.
-Get the indicated resin for
the photopolymerizer’s wave
length range, one with contains
photoinitiators based on
camphorquinone.
-Clean the light conductor.