22
English
English
Legal CE representative:
MPC International S.A
- 26, Boulevard Royal - L-2449 - Luxembourg - Fone: +49 431-7028 219 - Fax: +49 431-7028 202
Troubleshooting
Faults
Causes
Solutions
Not plugged in the power soure outlet
Plug it in
Equipment will not start
Fuse is burned out
Replace fuse
Power failure
Await normalization
Circuit breaker is off
Turn it on
System is not pressurized
Compressor is off and/or valve is closed
Turn compressor on and/or open valve
Tip
will not oscillate but LED goes on
Tip
is not correctly fixed
Fix tip appropriately or try another tip
Water leakage in the handpiece
Tip
is incorrectly screwed on
Screw tip on properly or replace O-ring
Sodium bicarbonate jet handpiece
Properly tighten the sodium bicarbonate
terminal is not properly screwed on
jet handpiece terminal
Tip
is clogged
Unclog
tip
Sodium bicarbonate will not come out
No sodium bicarbonate in the reservoir
Supply the reservoir
Sodium bicarbonate excess in the
Remove excess
Water will not come out
No irrigating solution in the reservoir
Supply the reservoir
Peristaltic pump is not enabled
Activate the peristaltic pump
Perception of electric shock on tip
No or ineffective grounding
Ground equipment adequately