MNL-00024 [1]
Proprietary and Confidential
Page
6
of
21
5.0
Transport and Storage Specifications
Ambient Temperature:
-40°C to 70°C (-40°F to 158°F)
Relative Humidity:
10% to 95%, Non-condensing
Atmospheric Pressure:
50 kPa to 106 kPa
Altitude:
≤ 2000m
6.0
EMC Precautions
The devices contained in this manual have been tested and found to comply with the EMC limits of the
international standard IEC 60601-1-2. These limits are designed to provide reasonable protection
against interference in a typical medical installation. The system can radiate radio frequency energy if
not installed in accordance with the instructions, and may cause harmful interference to other devices in
the vicinity. There is no guarantee that interference will not occur in a particular installation. If the
system does cause interference with other devices, which can be determined by turning the system off
and on, the user is encouraged to try to correct the interference by one or more of the following
measures:
•
Re-orient or relocate the receiving device
•
Increase the separation between the equipment
•
Connect the System into an outlet on a circuit different from that to which the other device(s) is
connected
Consult the manufacturer or field service technician for help
7.0
Electrical Specifications
The device is manufactured in the following electrical configurations. Always refer to the identification
label of the device to determine which electrical configuration the device is manufactured to.
•
Mains input = 230V, 3~, 60Hz, 100A Operating = 230V, 3~, phase to phase
•
Mains input = 400V, 3N~, 50Hz, 100A Operating = 230V~, single-phase (phase to neutral)
8.0
Site Selection
A site should be chosen within reach of the specified electrical connection (refer to the Facility
Requirements Guide) and where the unit can be left in place without obstructing traffic flow. It is
important that the unit be properly grounded. The site should not be in any area where water or
moisture might collect and should be protected from access by children and other unintended users.
WARNING:
To avoid the risk of electric shock this equipment must only be connected to a supply mains
with a protective earth.
CAUTION:
This device is a Class A Medical Device suitable for use in all establishments other than
domestic and those directly connected to the public low voltage power supply network that supplies
buildings used for domestic purposes.