![Daavlin ClearLink UV Series Operation Manual Download Page 5](http://html2.mh-extra.com/html/daavlin/clearlink-uv-series/clearlink-uv-series_operation-manual_5102551005.webp)
Company Confidential. Printed copies will not be updated
MNL-00002 [4]
Page
5
of
22
4.0
Site Selection
A location for the device should be chosen within reach of an electrical service that meets the
requirements of your specific device. See
Section
Error! Reference source not found.
below. Extension cords are not recommended. It is important that the unit be properly
grounded and installed on a standard, non-Residual Current Device (RCD) or ground fault circuit
interrupter (GFCI) breaker. It should
not
be in a location where there is foot traffic or where water
or moisture might collect. It should be protected from access by children. Exposure to ultraviolet
light over extended periods of time may cause carpets, wall coverings and furnishings to fade.
4.1
Electrical Requirements
The number of lamps in the unit determines the voltage, amperage, wire gauge, and the type of
electrical outlet necessary to safely support the unit’s operation. The table and diagrams below
define these requirements for each number of lamps.
Model
Amps
Wire Gauge
Diagram
Required Outlet
24 lamp, 120 volt
20 amps
12 AWG
A
NEMA L5-30R
16 lamp, 120 volt
15 amps
12 AWG
B
NEMA 5-15R
12 lamp, 120 volt
15 amps
14 AWG
B
NEMA 5-15R
Table 1 Electrical Requirements
4.1.1
Circuit Breaker
The unit’s circuit breaker is located under the frame, at the back right
-hand side of the device. Its location
is indicated by the circuit breaker label (right). When setting up the unit, ensure that the breaker is in the
ON
position.
5.0
Unpacking and Assembly
Note:
The UV Series rests on castors. If no stairs are involved, it may be easier to remove the
unit from its box at its delivery point (garage or foyer, for example) so it can be rolled into
position. If stairs are involved, you may consider hiring a professional moving company to
move the device into position.
1.
Using as many people as available, but a minimum of two, move the box (or the device, if
already unpacked) into the area where it will be used.
2.
To open, use metal cutters to cut the banding that encircles the shipping container.
B
MAIN DISCONNECT
CIRCUIT BREAKER
l
=
=
ON = RIGHT
O
=
=
OFF
=
LEFT