For example, each time you use the device, the ClearLink Controller will check the date of your 
last treatment.  If you are on schedule, it will automatically proceed.  However, if you have missed 
a treatment or treatments, the system will adjust accordingly to keep your dose at a safe level.  A 
screen will appear advising you of the adjustment (

See Figure 12 below

). 

 

Figure 12

 

 

Integrating dosimetry, another feature, uses a built-in sensor to monitor the amount of light energy 
being  produced  by  the  device.    The  system  then  automatically  adjusts  your  treatment  time  to 
account for variations in light energy caused by changes in room temperature or the aging of the 
lamps.    
Finally, before each treatment, the ClearLink Controller will ask about the current condition of 
your  skin,  such  as  whether  or  not  there  is  any  redness,  mild  or  otherwise.    It  will  use  this 
information to calculate the next dose.  Keep in mind, it is norma

l to experience some “redness” 

within 8 

 12 hours of a treatment but it is typically gone before it is time for the next treatment.   

11.2

 

Running a Treatment

 

Note: 

The  first  time  you  use  the  device,  your  dose  will  automatically  appear,  based  on  your 

doctor’s  prescription.  After  your  first  treatment,  the  screen  will  display  the  next  scheduled 

treatment, but will ask if you experienced any pain or redness after your last treatment.  Based on 
your response, the system will adjust the treatment according to your prescription. 
 

1.

 

After unlocking the device, the Treatment Screen will appear with the scheduled dose. 

2.

 

You will be asked, 

“Were you in pain or red after your last treatment?”  

 

3.

 

You must choose one of the following responses:

 

   

  No, or < 24 Hrs   

(“Not at all, or it was gone in less than 24 hours”)

 

   

  24 

 48 Hrs         

(“I was in pain or red for 24

-

48 hours after the treatment”)

 

   

  > 48 Hrs              

(“I was in pain or red for more than 48 hours”)

 

   

  

Still red / pain      (“I am currently still red or have pain”)

 

Note:

  

Selection of the “Still red / pain” option will lock the device until one treatment has been 

skipped.  This prevents treating the skin until the erythema (redness) has resolved.   

4.

 

Once you have highlighted your response, press the Enter   

   key.      

5.

 

The Treatment Confirmation screen will appear and will display the dose and an estimated 
time for the treatment. If a change is needed, press the back     arrow. To cancel press the 
Cancel     
         button. 

Summary of Contents for ClearLink UV Series

Page 1: ...Company Confidential Printed copies will not be updated MNL 00002 4 Page 1 of 22 ClearLink Home Control Operation Manual Modes of Operation Guided Mode Dosimetric Mode Timed Mode...

Page 2: ...2 11 1 Features of the ClearLinkTM Guided Home Control System 12 11 2 Running a Treatment 13 12 0 Dosimetric Mode 14 12 1 Features of the ClearLink Dosimetric Home Control System 14 12 2 Running a Tre...

Page 3: ...00002 4 Page 3 of 22 15 0 Environmental Specifications 18 16 0 Warranty 20 16 1 Limited Warranty Policy 20 16 2 Warranty Coverage 21 16 3 Customer Responsibility 21 16 4 Warranty Service 21 16 5 Disp...

Page 4: ...imed You control the length of each treatment based on instructions from your doctor See pages 16 and 17 for operating instructions that are specific to this mode of operation FlexRx Your physician ma...

Page 5: ...The table and diagrams below define these requirements for each number of lamps Model Amps Wire Gauge Diagram Required Outlet 24 lamp 120 volt 20 amps 12 AWG A NEMA L5 30R 16 lamp 120 volt 15 amps 12...

Page 6: ...variety of different lamps each having a different effect on the skin It is important to check that the proper lamps are installed Consult the Lamp Specification Guide shown below and if there is any...

Page 7: ...nly be connected to a supply main with a protective earth To prevent electric shock remove power to the device prior to cleaning and servicing To eliminate the risk of fire when replacing the fuse rep...

Page 8: ...damage The device control system display is susceptible to damage from excessive force Avoid excessive force to the control system to prevent damage All treatments must be administered under the dire...

Page 9: ...nitrous oxide 9 0 Labels and Symbols A removable label Figure 1 Unlock Label located near the keypad provides information on how to unlock the unit Figure 1 Unlock Label If the unit is equipped with...

Page 10: ...rmation Figure 3 Warning Label Figure 4 Identification Label The serial number and date of manufacture of the device are printed on an identification label Figure 4 Identification Label that is locate...

Page 11: ...ding upon your particular circumstances may change these directions Always follow your physician s instructions The ClearLink Controller will track your treatment data However you may want a notebook...

Page 12: ...ter a treatment you may protect any areas that feel sunburned with sunscreen for the next several treatments or until they appear normal Make a note in your notebook You should see your physician at t...

Page 13: ...t is time for the next treatment 11 2 Running a Treatment Note The first time you use the device your dose will automatically appear based on your doctor s prescription After your first treatment the...

Page 14: ...to monitor the amount of light energy being produced by the device The system then automatically adjusts the length of each treatment to account for variations in light output caused by changes in roo...

Page 15: ...ber and close the doors behind you Face either the back or front of the device In order to receive even and effective treatments center yourself within the unit As nearly as possible your torso should...

Page 16: ...e limiting software For each sequence of exposures the number of available treatments is reduced by one A treatment sequence consists of the primary exposure plus up to three repeat exposures if neces...

Page 17: ...ocol between the use of the device on different patients Use a high level disinfectant such as Steris Corporation s Resert XL HLD and follow the manufactures guidelines See also FDA Cleared Sterilants...

Page 18: ...or additional information Note that timely lamp replacement by Daavlin or its authorized agent will avoid this situation 1 First unplug the device then using a hex driver socket remove the screw locat...

Page 19: ...ctrical fast transient burst immunity test following IEC 61000 4 4 2kV 2kV Surge immunity test following IEC 61000 4 5 0 5 kV 1 0 kV 2kV 0 5 kV 1 0 kV 2kV Conducted Immunity test following IEC 61000 4...

Page 20: ...y occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 The...

Page 21: ...customer installation improper electrical service lack of proper maintenance and Acts of God This Limited Warranty does not extend to repairs made necessary by use of parts or accessories not recommen...

Page 22: ...declared null and void if non qualified technicians perform any repair or maintenance on the Equipment unless prior written authorization has been obtained from Daavlin Even with Daavlin s authorizat...

Reviews: