ClearLink
™
Home-Control
Operation Manual
Modes of Operation
•
Guided Mode
Dosimetric Mode
Timed Mode
Page 1: ...Company Confidential Printed copies will not be updated MNL 00002 4 Page 1 of 22 ClearLink Home Control Operation Manual Modes of Operation Guided Mode Dosimetric Mode Timed Mode...
Page 2: ...2 11 1 Features of the ClearLinkTM Guided Home Control System 12 11 2 Running a Treatment 13 12 0 Dosimetric Mode 14 12 1 Features of the ClearLink Dosimetric Home Control System 14 12 2 Running a Tre...
Page 3: ...00002 4 Page 3 of 22 15 0 Environmental Specifications 18 16 0 Warranty 20 16 1 Limited Warranty Policy 20 16 2 Warranty Coverage 21 16 3 Customer Responsibility 21 16 4 Warranty Service 21 16 5 Disp...
Page 4: ...imed You control the length of each treatment based on instructions from your doctor See pages 16 and 17 for operating instructions that are specific to this mode of operation FlexRx Your physician ma...
Page 5: ...The table and diagrams below define these requirements for each number of lamps Model Amps Wire Gauge Diagram Required Outlet 24 lamp 120 volt 20 amps 12 AWG A NEMA L5 30R 16 lamp 120 volt 15 amps 12...
Page 6: ...variety of different lamps each having a different effect on the skin It is important to check that the proper lamps are installed Consult the Lamp Specification Guide shown below and if there is any...
Page 7: ...nly be connected to a supply main with a protective earth To prevent electric shock remove power to the device prior to cleaning and servicing To eliminate the risk of fire when replacing the fuse rep...
Page 8: ...damage The device control system display is susceptible to damage from excessive force Avoid excessive force to the control system to prevent damage All treatments must be administered under the dire...
Page 9: ...nitrous oxide 9 0 Labels and Symbols A removable label Figure 1 Unlock Label located near the keypad provides information on how to unlock the unit Figure 1 Unlock Label If the unit is equipped with...
Page 10: ...rmation Figure 3 Warning Label Figure 4 Identification Label The serial number and date of manufacture of the device are printed on an identification label Figure 4 Identification Label that is locate...
Page 11: ...ding upon your particular circumstances may change these directions Always follow your physician s instructions The ClearLink Controller will track your treatment data However you may want a notebook...
Page 12: ...ter a treatment you may protect any areas that feel sunburned with sunscreen for the next several treatments or until they appear normal Make a note in your notebook You should see your physician at t...
Page 13: ...t is time for the next treatment 11 2 Running a Treatment Note The first time you use the device your dose will automatically appear based on your doctor s prescription After your first treatment the...
Page 14: ...to monitor the amount of light energy being produced by the device The system then automatically adjusts the length of each treatment to account for variations in light output caused by changes in roo...
Page 15: ...ber and close the doors behind you Face either the back or front of the device In order to receive even and effective treatments center yourself within the unit As nearly as possible your torso should...
Page 16: ...e limiting software For each sequence of exposures the number of available treatments is reduced by one A treatment sequence consists of the primary exposure plus up to three repeat exposures if neces...
Page 17: ...ocol between the use of the device on different patients Use a high level disinfectant such as Steris Corporation s Resert XL HLD and follow the manufactures guidelines See also FDA Cleared Sterilants...
Page 18: ...or additional information Note that timely lamp replacement by Daavlin or its authorized agent will avoid this situation 1 First unplug the device then using a hex driver socket remove the screw locat...
Page 19: ...ctrical fast transient burst immunity test following IEC 61000 4 4 2kV 2kV Surge immunity test following IEC 61000 4 5 0 5 kV 1 0 kV 2kV 0 5 kV 1 0 kV 2kV Conducted Immunity test following IEC 61000 4...
Page 20: ...y occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 The...
Page 21: ...customer installation improper electrical service lack of proper maintenance and Acts of God This Limited Warranty does not extend to repairs made necessary by use of parts or accessories not recommen...
Page 22: ...declared null and void if non qualified technicians perform any repair or maintenance on the Equipment unless prior written authorization has been obtained from Daavlin Even with Daavlin s authorizat...