4.2  External Detector Module 

The  External  Detector  module  houses  a  0.5”  diameter  x  1.5”  Cesium  Iodide  Scintillation  crystal  with a 
high-sensitivity photo-multiplier tube.  It detects gamma radiation within the energy range of 30 keV to 3 
MeV.  It takes readings in less than one second. 

4.3  Protective Shield 

Included with the G2 model, the protective shield adds an increased level of environmental protection by 
prohibiting the buildup of snow and ice on the cable and connector on the External Detector module.  The 
protective shield comes attached to the External Detector module. 

4.4  Cables 

There are 2 cables included with the Mini Rad-V – the Power cable and the Communication cable.  The 
Power cable connect the In-cab Display module to the vehicle power.  The Communication cable connects 
the In-Cab Display module to the External Detector module. 

The standard cable lengths for both cables are 20 feet long.  Longer cables lengths of 35’ and 50’ are offered 
as options. 

4.5  Maintenance 

The External Detector module is designed to MIL-STD-810 environmental standards for rain, shock, and 
vibration. The unit is sealed to reduce intrusion from dust, humidity, and salt fog. It may be cleaned safely 
with mild soap and water using a warm, wet cloth.

 

Installation Instructions 

Please  refer  to  your  vehicle  manufacturer’s  manual  for  specific  installation  guidance  for  your  vehicle, 
particularly for mechanical mounting of the Mini Rad-V modules, routing of the power and communication 
cables,  and  connecting  the  Mini  Rad-V  to  vehicle  power.  D-tect  Systems  assumes  no  responsibility  for 
damages to any vehicle, other electronic devices, or this product incurred during installation. 

5.1  Installation Steps 

Warning: 

Do  not  apply  power  to  the  In-Cab  Display  module  until  the  External  Detector  module  is 

connected to the In-Cab Display module with communication cable. 

The In-Cab Display module must be connected to the External Detector module BEFORE power is applied, 
and  the  In-Cab  Display  module  should  NEVER  be  connected  to  power  without  the  External  Detector 
module attached. 

Please follow these instructions exactly.  Failure to do so may damage the unit.   

1.

 

Mount  the  External  Detector  module  to  the  bumper  or  other  location  as  appropriate.    Do  NOT 
connect the communication cable to the External Detector module at this time. 

2.

 

Mount the In-Cab Display module onto the dashboard of the vehicle. Do NOT connect the detector 
cable to the External Detector module at this time. 

3.

 

Run  the  communication  cable  from  the  In-Cab  Display  module  to  the  to  the  External  Detector 
module. Do not connect the cable to the External Detector module.  Secure the cable to the bumper 
with a zip tie. While running the cables through the vehicle, make sure to not twist the connector 

Summary of Contents for Mini Rad-V

Page 1: ...13 D tect Systems 11814 South Election Rd Suite 200 Draper UT 84020 801 260 4000 INSTALLATION AND OPERATING INSTRUCTIONS FOR THE VEHICLE MOUNTED RADIATION DETECTION SYSTEM Generation 2 Version 5 Augus...

Page 2: ...s been included to mitigate the collection of snow ice and moisture on the Exterior Detector housing The entire installation process was re evaluated as well and is now simpler and more straightforwar...

Page 3: ...structions Date 8 28 2019 Page 3 of 13 D tect Systems 11814 South Election Rd Suite 200 Draper UT 84020 801 260 4000 Figure 1 In Cab Display Module Figure 2 External Detector Module Figure 4 External...

Page 4: ...afely with mild soap and water using a warm wet cloth 5 Installation Instructions Please refer to your vehicle manufacturer s manual for specific installation guidance for your vehicle particularly fo...

Page 5: ...ached to the vehicle with four screws spaced 2 apart on both ends of the module The sensitivity of the device decreases with the amount of material between the detector and any potential radiation sou...

Page 6: ...of the driver or interfere with vehicle controls and or other equipment Contact your local Department of Transportation for information regarding regulations for dashboard mounted displays as these m...

Page 7: ...nstalling Cables The modules are connected by the communication cable included To connect the cable adjust it until it fits onto the post then turn the cable connector until it locks firmly into the p...

Page 8: ...the blue LED will flash to indicate the 30 second calibration phase When calibration is complete the Mini Rad V will beep once more and go into detection mode This keeps the Mini Rad V at maximum sens...

Page 9: ...viewing Alarm Loud audible alarm with acknowledge button Alarm Volume 90 dB at 10 cm Power 12V DC Energy Range 30 keV 3 MeV Environment In Cab Display Controlled environment operates in high RF envir...

Page 10: ...radiation levels it will recalibrate N A While the blue light is on the red light begins to flash and the audio begins to beep This is the expected behavior when detecting the presence of gamma and x...

Page 11: ...place or refund the purchase price of any Products that do not conform with this warranty Repair may include the replacement of parts with functionally equivalent reconditioned or new parts Replacemen...

Page 12: ...PURCHASE PRICE OF THE PRODUCT OR FOR ANY INDIRECT INCIDENTAL SPECIAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OR INABILITY TO USE THE PRODUCT TO THE FULL EXTENT THESE DAMAGES MAY BE DISCLAIMED...

Page 13: ...rts of them Laws in the United States and other countries preserve for D tect Systems certain exclusive rights for copyrighted D tect Systems software such as the exclusive rights to reproduce in copi...

Reviews: