Visual inspection: 

Only one master drive is allowed.
The master always has the last address of all the drives; this can be found on a sticker attached to the 
drive. Slaves are numbered downwards from the master, so that in a group of three drives, for example, 
the master is no. 3, with a slave 2 and slave 1 drive. Please note that this only applies in the ex-factory 
configuration. As soon as the drives are reconfigured with SCS, the ex-factory addressing no longer 
applies.

Wiring:

Is the drive or drive group wired properly?
See connection plans.

Diagnostics with SCS software:

The SCS software is required for further troubleshooting or configuration.

Troubleshooting

Minimum sash height/ -width

The details apply to a  

non-rotational installation of the drive!

Standard chain:

Minimum size (mm) = Stroke (mm) x 3.9

Side bow chain:

Stroke 

 

Minimum size (x)

400 mm  =  650 mm
500 mm  =  800 mm
600 mm  =  950 mm
800 mm  =  1250 mm
1000 mm  =  1550 mm

x

x

-  The drive set is fully configured ex factory.The 

drives must be reconfigured using SCS in the 
event of changes to the ex-factory configuration 
(e.g. when removing a drive from existing set or 
integrating a new drive).

-  Mount the drives whilst under zero stress. With 

that, an even load distribution is provided by all 
the drive chains.

-  At vertical active pulling application (e.g. for a 

skylight) the forces must be rearranged by 
software SCS.

-  If after mounting, the distance X is greater than 23 

mm, null balance with software SCS is neccesary. 
Closure  Ranges  will  shift  in  relation  to  new  zero 
point.

Mounting information

≥ 23 mm ? 

X

ACB bus interface

ACB is used for secure communication between the 
drive  and  compatible  D+H  control  systems.  It 
enables  activation,  diagnostics  and  configuration 
with perfect position, directly from the control panel. 
In  this  process,  all  status  messages,  such  as  the 
OPEN  and  CLOSED  signals,  opening  stroke  and 
drive errors, are transmitted to the control panel.

BSY+ synchronised electronics:

A  synchronous  group  can  contain  up  to  8  drives, 
which communicate via bus. Each drive has its own 
address, configurable with SCS.
The last drive of a synchronous group is the master, 
which  controls  the  other  drives,  the  slaves. 
Differences in forces between the drives of a group 
are  balanced  by  means  of  intelligent  force  and 
position  control.  In  the  event  that  a  drive 
malfunctions, all the drives are automatically cut off.

Functional description

Main closing edge

Beside closing edge

Beside
closing edge

In  the  "CLOSE"  direction  the  drive  has  an  active 
protection for the main closing edge. If there is an 
overload in the closing ranges C3 and C2, the drive 
runs "OPEN" for 10 seconds, then "CLOSE" again. If 
closing is not possible after three attempts, the drive 
remains in this position.
In  addition,  the  drive  is  equipped  with  passive 
protection.  The  closing  speed  in  closing  range  C2 
and C1 is reduced to 5 mm/s.

 

  Significantly greater forces can be

 

  exerted at the secondary closing

 

  edges. Danger of crushing hands and

 

  fingers!

Entrapment protection

 (see N.12 of Safety Notes; p 2/20)

Summary of Contents for CDC-0252-1-ACB

Page 1: ...por separado Standard version R Ex cution standard R Est ndar modelo R Mirrored version option L Ex cution invers e option L Versi n invertida Opci n L en Original instructions Page 2 Connection Page...

Page 2: ...after closing Direct connection of additional closing edge protection at the drive possible SKS option WARNING Read all safety warnings instructions illustrations and specifications provided with this...

Page 3: ...lity Ambient humidity Emission sound pressure level Additional functions approx 1500 N without console 10 000 double strokes 30 S6 EN 60034 1 Aluminium powdercoated RAL 9006 IP 32 15 75 C 30 min 300 C...

Page 4: ...ng stroke and drive errors are transmitted to the control panel BSY synchronised electronics A synchronous group can contain up to 8 drives which communicate via bus Each drive has its own address con...

Page 5: ...nly applies in the ex factory configuration As soon as the drives are reconfigured with SCS the ex factory addressing no longer applies Wiring Is the drive or drive group wired properly See connection...

Page 6: ...nd cleaning Maintenance work is only allowed when the device is in a de energized condition Inspection and maintenance has to be carried out according to D H maintenance notes Only original D H spare...

Page 7: ...ns Tenirlespersonnes l cartdelazone demouvementdelamotorisation Tenirlesenfants l cartdelacommande Respecterlediagrammedechargedepression delacha ne N utiliserquedansdeslocauxsecs Uniquementpourmontag...

Page 8: ...pure En peu de temps C3 250 mm 250 N Plages d ouverture et de fermeture O1 C1 C2 O2 O3 23 mm 150 N 77 mm 150 N 23 mm 250 N 15 mm 250 N 250 N Type CDC 0252 0350 1 ACB CDC 0252 0500 1 ACB CDC 0252 0600...

Page 9: ...pe de synchronisation peut se composer de maximum 8 moteurs qui communiquent par le biais d un bus Chaque moteur poss de sa propre adresse laquelleestconfigurable l aidedulogicielSCS Le dernier moteur...

Page 10: ...l tat la livraison En effet en cas de reconfiguration avec le logiciel SCS l adressage d fini en usine n est plus valable C blage Le moteur ou le groupe de moteurs est il correctement c bl Voir plans...

Page 11: ...vice avait t entrepris e par un partenaire de service et de vente D H autoris Si vous raccordez les composants D H des installations trang res ou panachez les produits D H des pi ces d autres construc...

Page 12: ...ficharojadeseguridadadjunta 5 Peligrodeatrapamientoconfuertepresi nenlas reasaccesiblesconlasmanos 6 Los operadores de las ventanas han sido evaluados nicamente para su uso residencial o para el uso c...

Page 13: ...ca de emisi n Funciones adicionales aprox 1500 N sin fijaci n 10 000 dobles carreras 30 S6 EN 60034 1 Aluminio Pintura en polvo RAL 9006 IP 32 15 75 C 30 min 300 C 90 sin condensaci n LpA 35 dB A Acti...

Page 14: ...nizaci n Un grupo sincronizado puede estar formado por un total de hasta 8 motores que se comunican a trav s de un bus Cada motor tiene su propia direcci n la cual puede configurarse mediante el softw...

Page 15: ...e vuelvan a configurar los motores con el software SCS las direcciones de f brica ya no ser n aplicables Cableado Se ha realizado correctamente el cableado del accionamiento o del grupo de accionamien...

Page 16: ...echadelaentrega siemprequeelmontaje y la puesta en marcha hayan sido realizados por un distributor autorizado de D H La garant a de D H expira en el caso de conexi n de componentes de D H con otros eq...

Page 17: ...erde PK pink rose rosa GY grey gris Option SKS Opci n SKS BN Mot b OG SKS GY SKS GN BSY b YE BSY a WH Mot a BN Mot b OG ACB a HS GY ACB b GN BSY b YE BSY a WH Mot a Option SA SZ Opci n SA SZ BN Mot b...

Page 18: ...MOT BK BK OG 2x47kW End mod ul Term inal mod ule Mod ule term inal EM 47 K WH BN OG YE GN GY Mot a Mot b ACB a HS BSY a BSY b ACB b CDC 0252 1 Junction box Bo te de d rivation Caja de conexi n Master...

Page 19: ...k rose rosa GY grey gris gris Monitoring Supervision Supervisi n Not in case of Pas pour No en el caso GVL E K M Protect wire against short circuit Prot ger contre les courts circuits Proteger contra...

Page 20: ...mersbek Rights to technical modifications reserved Sous r serve de modifications techniques Derecho reservado a modificaciones t cnicas 100 Recycled paper Papier recycl Papel reciclado B C 30 mm 3 5 m...