![cytiva Sterile Tube Fuser Operating Instructions Manual Download Page 48](http://html1.mh-extra.com/html/cytiva/sterile-tube-fuser/sterile-tube-fuser_operating-instructions-manual_2706957048.webp)
Insert a cutting blade
CAUTION
The cutting blade of Sterile Tube Fuser - Dry is a thin piece of stain-
less steel. It does not have a sharpened edge but can cause cuts if
handled improperly. Be sure to read and understand the procedure
of how to install and remove the blade.
Step
Action
1
Place the Sterile Tube Fuser - Dry blade in Installation/Removal Tool as
shown below.
2
Use Installation/Removal Tool to install the blade by pressing the top of the
tool with the palm of the hand. During the blade installation, secure the
blade within the tool by compressing with the thumb as shown in the image
below.
3
Press Installation/Removal Tool with the palm of the hand until the blade is
fully inserted into mounted position. Remove the Installation/Removal Tool.
5 Operation
5.2 Sterile Tube Fuser - Dry
5.2.2 Fusing tubes
48
Sterile Tube Fuser - Dry and Hot Lips Tube Sealer Operating Instructions 28964139 AG