cyclotest mySense Instructions For Use Manual Download Page 8

EN-14

EN-15

Operating the device

Operating the device

•  Unlock the battery cover by turn-

ing it past the marker to the left. 

Then lift the cover off.

•  Remove the old battery and insert 

a new battery (CR2032) with the 

“+” symbol facing up.

•  Place the cover back on the bat-

tery compartment and lock it in place.

Battery display/changing the battery

The device is equipped with a CR2032 lithium battery.

If the battery icon   flashes on the display while 

the thermometer is in use, the battery is low 

(Figure 1). You can still perform some tempera-

ture measurements at this point, but you should 

have a spare battery ready.

After changing the battery, the thermometer must be connected to 

the mySense app so that the date and time can be resynchronised. 

Until you have done this, you cannot take any new temperature meas-

urements.

Change the battery if:

–  After switching on the device, the “battery 

flat” indicator (Figure 2) appears on the dis-

play. You must insert a new battery before 

you can take any temperature measure-

ments

–  Nothing is shown on the display despite 

pressing the on/off button.

Figure 1

Figure 2

Battery 

compartment 

cover locked

Battery com-

partment cov-

er unlocked

Measurement values stored on the device are retained when the bat-

tery is changed.

Changing devices

Each mySense basal body temperature thermometer has a unique 

signature assigned to a specific user. Signatures are assigned during 

initial setup when the device is synchronised with the mySense app 

for the first time.

If your device is lost or faulty, you can assign a new mySense basal 

body temperature thermometer to your user account.

To do this, follow the instructions in the chapter “Initial setup” on page 7.

Please note: If you change device during a cycle, the current cycle be-

comes inactive in the mySense app. For your safety, the mySense app 

only accepts measurement data from a single device during a cycle.

Summary of Contents for mySense

Page 1: ...Instructions for use www cyclotest de mySense...

Page 2: ...anging the battery Changing devices Error messages Technical data Customer service Explanation of symbols Disposal 3 4 6 7 7 9 12 13 14 15 16 17 18 19 19 20 21 22 The cyclotest mySense is a medical di...

Page 3: ...d measuring tip can cause injuries Do not bite or bend the sensor This can cause damage and lead to malfunctions The use of this device near mobile phones radios or other strong electromagnetic fields...

Page 4: ...Bluetooth data connection Temperature unit C F Synchronising Battery display Time Date Memory mode Initial device setup You must install the cyclotest mySense app on your smartphone be fore setting u...

Page 5: ...he device cannot connect to the app it switch es off after 60 seconds If this occurs check whether Bluetooth is enabled on your smart phone and that the mySense app is running Now try to establish a c...

Page 6: ...es a temperature that is higher than 37 50 C the display switch es between the temperature and HI fever warning Figure 10 This measurement will not be included in the cyclotest mySense app s cycle cal...

Page 7: ...edly pressing the memory recall key you can call up the previous measure ments no 2 no 3 If no measurement values are saved the fol lowing appears on the display Figure 4 Figure 3 Figure 2 Figure 1 Fi...

Page 8: ...r switching on the device the battery flat indicator Figure 2 appears on the dis play You must insert a new battery before you can take any temperature measure ments Nothing is shown on the display de...

Page 9: ...ry ready Error messages No display after switching on Check whether the bat tery is positioned correctly and the polarity is cor rect If the error persists change the battery Bluetooth connection erro...

Page 10: ...humidity 15 to 95 Battery 1 x CR2032 3 0 V mercury free Battery life approx 400 measurements Data transfer Bluetooth 4 0 LE low energy IP classification IP67 dust tight protection against temporary su...

Page 11: ...our expense in our plant through repair work or replacement delivery of a defect free device The warranty does not cover normal wear and tear of wearing parts or damage caused by failure to observe t...

Page 12: ...haracter istics If you have any additional questions you can call our cyclotest ser vice hotline which is available Monday to Thursday from 8 00am to 5 00pm and Fridays until 4 30pm Your cyclotest adv...

Page 13: ...odifications Reprinting including of extracts is prohibited Copyright 2018 UEBE Medical GmbH UEBE Medical GmbH Zum Ottersberg 9 97877 Wertheim Germany Phone 49 0 93 42 92 40 40 Fax 49 0 93 42 92 40 80...