![Cybertek EM5080 Series Manual Download Page 14](http://html1.mh-extra.com/html/cybertek/em5080-series/em5080-series_manual_2703117014.webp)
Shenzhen Zhiyong Electronics Co., Ltd
13
Draw and call the compensation curve
①
Draw compensation curve
Press
to enter the main menu
Select Set---Compensate to enter compensation edit interface
Take our LISN EM5040B as example. Due to the inner 10dB attenuator, the actual test result need
compensate for 10dB. Type in corresponding frequency in Freq, enter compensation value 10 in Factor
and select Add, Delete or Reset according to your need. Name the compensation curve LISN-COMP
in the empty space of Filename and click on Save or Save To Default.
②
Load Compensation Curve
Click on Set-Compensation Antenna-Select, select the compensation curve needed. Click on Use
Compensation, and the program will apply the current compensation curve and display the compensation
value on the sheet. Click on OK to confirm.