1. Summary 

EM5018 Absorbing Clamp is mainly used for measuring the power interference signal from the 

power cable of EUT, frequency ranging from 30MHz to 1000MHz. EM5018 is in accordance with 

CISPR  16-1-3  regulation,  with  operating  current  over  25A,  pore  diameter  around  20mm,  can  be 

widely used in the measurement of testing the interference signal of the standard power supply cable 

like CISPR 13/EN 55013, CISPR 14-1/EN 55014-1.   

2.

 

Product and Accessories   

 

 

Gasket Clamp: can increase the measurement accuracy when the diameter of cable under test 

is less than 6mm. If the diameter is over 6mm, remove the gasket clamp.   

 

EUT port: used to connect the cable of EUT. 

 

Spring catch: locked during measurement 

 

RF  output  port:  interference  signal  output  port,  connecting  with  measuring  devices  like 

receiver. 

3.

 

Electronics Specification 

Frequency range 

30 MHz~1000MHz 

Input loss 

13 dB to 22 dB 

Clamp factor 

-4 dB to +5 dB    refer to the diagram below 

Maximum allowable disturbance power 

<10 W 

Maximum allowable   

cable diameter 

 

20 mm 

Clamp with gasket 

6mm 

Maximum allowable DC or AC peak   

working current 

25 A 

RF Output port 

BNC, 50 Ω 

Operating temperature 

+4°C to +40°C 

Dimension

W×H×D

mm 

around 648*105*85mm 

Weight 

around 6.5kg 