Typical application configuration diagram of power disturbance
measurement
When the diameter of the cable under test is over 6mm, remove the gasket clamp of the
absorbing clamp, which can increase precision when the cable is thin
Important tips during measurement:
Please lock the absorbing clamp.
The side with EUT mark should be close to the EUT, absorbing clamp cannot be installed
reversely.
Find maximum value of all frequency point by moving absorbing clamp on the trail.
Interference signal measured P (dB(pW)) = Voltage measured V(dB(uV)) + Clamp factor
(dB(pW/uV))