Typical  application  configuration  diagram  of  power  disturbance 

measurement 

 

 

 

When  the  diameter  of  the  cable  under  test  is  over  6mm,  remove  the  gasket  clamp  of  the 

absorbing clamp, which can increase precision when the cable is thin 

 

Important tips during measurement: 

 

Please lock the absorbing clamp. 

 

The side with EUT mark should be close to the EUT, absorbing clamp cannot be installed 

reversely.   

 

Find maximum value of all frequency point by moving absorbing clamp on the trail. 

 

Interference  signal  measured  P  (dB(pW))  =  Voltage  measured  V(dB(uV))  +  Clamp  factor 

(dB(pW/uV))

 