connect the BNC output port of the probe and input terminal of the oscilloscope using the standard 

BNC cable. 

 

Turn on the voltage switch, the indication light should be lighted green. 

 

Press the button to select the proper range according to the value of the current under test. 

 

Attention: Different range corresponds to different attenuation ratio   

 

Press the auto zero button to realize zero set. The buzzer will buzz shortly for twice if auto zero set 

successfully, or it will buzz for once, meaning auto zero set fails. One thing need attention is that, the 

external magnetic field could have slight effect on the AC zero position of the probe, so user better 

not move the probe after zero set complete.   

 

Open the clamp of the current probe and fix the conductor under test 

Attention:  there’s  direction  indication  on  the  clamp,  the  same  current  direction  will  generate  positive 

output, and otherwise it will generate negative output. 

 

Properly  set  the  vertical  sensitivity  of  the  oscilloscope  to  generate  stable  wave  form.  When 

oscilloscope is set DC CM, user can observe both DC and AC component, but if there’s only AC CM, 

only AC component will be seen. 

5.Mechanic Specification 

Size of current clamp 

        216*115*45mm 

Size of output box 

        137*33*35mm 

Operating altitude

 

    0~2000m

 

Max size of conductor under test

 

        diameter 54mm

 

Length of the connecting cable between   

clamp and box 

        1m 

Length of double-terminal BNC cable

 

        1m

 

Weight

 

        620g (Battery Excluded)

 

 

6.Environmental Specification

 

Operating temperature

 

        0℃

+50℃

 

Storing temperature

 

-20℃

+80℃

 

Operating relative humidity

 

        0℃~+40℃

Humidity 95%RH

 

+40℃~+50℃

Humidity 45%RH

 

Pollution level

 

Level 3

 

 

7.Maintenance   

In the product warranty period, if the product is used normally, the malfunction is caused 

by  the  quality  problem  and  is  not  dismantled,  our  company  will  provide  free  repairing 

service.   

 

Clamp: the clamp should be clean and dry. If there’s a need of cleaning up, please use soft cloth 

together with alcohol to erase the dirt. Do not put the clamp in wet environment. 

 

Hand  shank:  Please  clean  up  the  hand  shank  using  clean  cloth  or  sponge.  Do  not  use  water  to 

CYBERTEK

Summary of Contents for CPL2000

Page 1: ...Shenzhen Zhiyong Electronics Co Ltd www cybertek cn AC DC Low Frequency Current Probe CPL2000 2000A 10kHz Shenzhen Zhiyong Electronics Co Ltd CYBERTEK...

Page 2: ...ut voltage surpassing the safety range if the basic insulation of the test terminal does not meet the requirement Please refer to the matters need attention about electronic safety Electric shock coul...

Page 3: ...has an inner square shape 9V battery CPL2000 is highly power consuming so please use the alkaline battery that can last long Comparatively large test error could occur when the power supply is low Wh...

Page 4: ...eface 1 CPL2000 Brief Description 2 Summary 4 Probe Parts Description 4 Electronics Specification 5 Operating Method 5 Mechanic Specification 6 Environmental Specification 6 Maintenance 6 Treatment wh...

Page 5: ...s it to connect with oscilloscope and other devices including multi meter by BNC to Double Banana plug CPL2000 is usually used in fields including power frequency measurement motor drive and power sup...

Page 6: ...Peak value current 300Apk 3000Apk Range Sensitivity 10mV A 1mV A Typical Accuracy DC 45Hz 66Hz 2 rdg 5mV 2 rdg 1mV Bandwidth 3dB DC 10kHz Operating voltage CATII 600V Typical battery type and service...

Page 7: ...t Properly set the vertical sensitivity of the oscilloscope to generate stable wave form When oscilloscope is set DC CM user can observe both DC and AC component but if there s only AC CM only AC comp...

Page 8: ...tery voltage lower than 6 5V Change the battery Amplitude low in all frequency range The input resistance of other testing device is 50 Please set the input resistance over 1M 9 Packing List PACKING L...