clean the equipment, use alcohol instead and dry it after the treatment. 

 

To guarantee the performance, user can proceed check or calibration once a year. 

8.Treatment when irregular occurs 

Problem 

Possible Reason 

Treatment 

Unable to test DC, or amplitude   

too low 

Power supply is off 

Turn on the power supply 

Oscilloscope is set AC CM 

Please set to DC CM 

Clamp is not completely closed 

Close the clamp completely 

Power Indication Light is not   

lighted after device is turned on 

Battery voltage lower than 6.5V 

Change the battery 

Amplitude low in all frequency range 

The input resistance of other 

testing device is 50Ω 

Please set the input resistance 

over 1MΩ 

 

9.Packing List 

PACKING LIST 

Name 

Number 

Current probe 

9V battery 

DC12V/1A adapter

CK-612) 

BNC output cable (CK-310) 

High class tool kit 

Instruction manual 

Warranty card 

Test report 

 

 

 

 

CYBERTEK   

     

SHENZHEN ZHIYONG ELECTRONICS CO., LTD

 

Addr

 

Room A1702, Building 4, TianAn Cyber Park, HuangGe North Road, LongGang                                           

District, ShenZhen City, China 

Tel:     

+86-400 852 0005 / +86-755-86628000   

Q Q:   

400 852 0005 

Fax: 

  +86)0755-8662 0008   

Email

[email protected]

                                               

© Zhiyong Electronics, 2018

 

Url: 

    http://www.cybertek.cn                                               

Published in China, January 1, 2018 

 

CYBERTEK

Summary of Contents for CPL2000

Page 1: ...Shenzhen Zhiyong Electronics Co Ltd www cybertek cn AC DC Low Frequency Current Probe CPL2000 2000A 10kHz Shenzhen Zhiyong Electronics Co Ltd CYBERTEK...

Page 2: ...ut voltage surpassing the safety range if the basic insulation of the test terminal does not meet the requirement Please refer to the matters need attention about electronic safety Electric shock coul...

Page 3: ...has an inner square shape 9V battery CPL2000 is highly power consuming so please use the alkaline battery that can last long Comparatively large test error could occur when the power supply is low Wh...

Page 4: ...eface 1 CPL2000 Brief Description 2 Summary 4 Probe Parts Description 4 Electronics Specification 5 Operating Method 5 Mechanic Specification 6 Environmental Specification 6 Maintenance 6 Treatment wh...

Page 5: ...s it to connect with oscilloscope and other devices including multi meter by BNC to Double Banana plug CPL2000 is usually used in fields including power frequency measurement motor drive and power sup...

Page 6: ...Peak value current 300Apk 3000Apk Range Sensitivity 10mV A 1mV A Typical Accuracy DC 45Hz 66Hz 2 rdg 5mV 2 rdg 1mV Bandwidth 3dB DC 10kHz Operating voltage CATII 600V Typical battery type and service...

Page 7: ...t Properly set the vertical sensitivity of the oscilloscope to generate stable wave form When oscilloscope is set DC CM user can observe both DC and AC component but if there s only AC CM only AC comp...

Page 8: ...tery voltage lower than 6 5V Change the battery Amplitude low in all frequency range The input resistance of other testing device is 50 Please set the input resistance over 1M 9 Packing List PACKING L...

Reviews: