63
Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13
R
ECOMMENDED
SEPARATION
DISTANCES
BETWEEN
PORTABLE
AND
MOBILE
RF
COMMUNICATIONS
EQUIP
-
MENT
AND
THE
EC
5000
ERGOMETER
The ec5000 ergometer is intended for use in an electromagnetic environment, as specifi ed below, in which radiated RF disturbances are controlled.
The customer or the user of the ec5000 ergometer can help prevent electromagnetic interference by maintaining a minimum distance between por-
table and mobile RF communications equipment (transmitters) and the ec5000 ergometer as recommended below, according to the maximum output
power of the communications equipment
Rated Maximum Out-
put Power of Transmit-
ter [W]
separation distance according to frequency of transmitter
[m]
150 kHz to 80 MHz
d = 1.2 √P
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2.5 GHz
d = 2.3 √P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.7
3.7
7.37
100
11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection from
structures, objects, and people.
Summary of Contents for Ergometer custo ec5000
Page 1: ...EXT 0001 DK 1585 Version 001 2015 11 13 EL 20100043500 Operating Manual Ergometer custo ec5000...
Page 4: ...4 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 6: ...6 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 34: ...34 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 54: ...54 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...