43
Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13
Protocols
0. WHO
1. BAL
2. Hollm
3. Std Fr
4. Standard
5. Programm 6
6. Programm 7
7. Programm 8
8. Programm 9
9. Programm 10
Description
Programm 6
Edit
Type
Step
Basic Load:
30 W
Stage Time:
2 min
Stage Rate:
25 W
Recovery Time:
2 min
Recovery Load:
25 W
Description
Programm 6
Type
Step
2.
Basic Load
30 W
Stage Time
2 min
Stage Rate
25
W
P
ROTOCOLS
The first five exercise test protocols (WHO, BAL, Hollm, Std
Fr and Standard) are preconfigured, but all protocols in the
list are editable.
To reach the level for editing of the individual protocol pa-
rameters, first touch the protocol that you want to modify
(e.g., [ 5. Program 6 ]), and then touch the [ Edit ] key.
To change the name of a protocol, touch the name and
enter the new name from the keypad. Confirm your inputs
with the [ ] key.
At Type, you can choose [ Step ], [ Ramp ] or [ Inactive ].
You scroll through the Type options with the [ ] key.
When choosing the Step type (load increase in steps),
define the basic load (from 6 to 100 W), the stage time
(from 1 to 30 min) and the increment (from 1 to 400 W).
When choosing the Ramp type (continuous load increase),
define the basic load (from 6 to 100 W) and the load
increase (from 1 to 50 W).
P
ROTOCOL
CONFIGURATION
Summary of Contents for Ergometer custo ec5000
Page 1: ...EXT 0001 DK 1585 Version 001 2015 11 13 EL 20100043500 Operating Manual Ergometer custo ec5000...
Page 4: ...4 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 6: ...6 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 34: ...34 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...
Page 54: ...54 Ergometer custo ec5000 EXT 0001 DK 1585 Version 001 2015 11 13...