0 to the setting level, and from the setting value to 0. When the
ramp time is set, each contraction may be ramped up and down in
order that the signals come on and come off gradually and smoothly.
The ramp time is adjustable from 1 to 6 seconds.
7) Set Contract Time
The contract Time controls the time of stimulation and can be
adjusted. Press [S] button cycle to enter this menu, and then press
the [
▲
] and [
▼
] button to adjust the setting. Both channels’
stimulation is cycled on and off by the contraction and relaxation
settings. The range is adjustable from 1 to 60 seconds.
Caution:
Contract time does not include the ramp up and ramp down
time, ON time=Ramp up + Contract time + Ramp down.
8) Set Relaxation (OFF) time
The Off Time controls the time of relaxation and can be adjusted.
Press [S] button cycle to enter this menu, and then press the [
▲
]
and [
▼
] button to adjusting the setting. Both channels’ stimulation
is cycled on and off by the contraction and relaxation settings. The
range is adjustable from 0 to 60 seconds.
In Alternate program, the OFF Time should be equal or more than
the ON Time. (OFF TIME
≥
ON TIME)
4.7.3 IF Setting
Press the [S] button to enter the setting state. The settings can be
adjusted according to the following steps:
1) Set the Therapeutic Program
There are 4 programs in IF therapeutic mode
available. The therapeutic program can be
selected by pressing the [
▲
] and [
▼
] button.
The mode you selected will show up on the
top of liquid crystal display.
2) Set Timer
Press [S] button cycle to enter this setting.The treatment time is
adjustable from 1 to 60 minutes or Continuous.
29
Summary of Contents for InTENSity Select Combo
Page 1: ...T E N S P r o s SELECT combo INSTRUCTION MANUAL ...
Page 9: ...T E N S P r o s 1 5 Warnings Cautions and Adverse Reactions WARNINGS 9 ...
Page 10: ...T E N S P r o s ...
Page 16: ...T E N S P r o s ...
Page 22: ...T E N S P r o s Synchronous S Alternate A Delay D Interferential Microcurrent Constant 22 ...
Page 43: ...T E N S P r o s ...