TENSPros
3. Specification
3.1 Accessories
3.2 Technical information
Channel
Dual, isolated between channels
Power supply
9.0 V Alkaline DC -1 *6LR61 battery
Adapter output: 9.0Vdc, 800mA(optional)
Operating conditions
5°C to 40°C
(
41
℉
to 104
℉)
with a relative
humidity of
Storage conditions
-10°C to 50°C
(
14
℉
to 122
℉)
with a relative
humidity of
Dimensions
4.5×2.55×0.9 inches(L*W*H)
Weight
0.28 lbs(With battery)
Tolerance
There may be a ±5% tolerance of all setting
and ±10%
Electrode Detection
Function
The amplitude level will be reset to 0mA when the
amplitude level is 12mA or greater and an open
circuit at either channel is detected.
Timer
Adjustable from 1 to 60 minutes or continuous.
Adjusts in 1 minute steps. Treatment time
countdown is automatic.