![Curative CS-HYB-MO Instructions Manual And User Manual Download Page 8](http://html1.mh-extra.com/html/curative/cs-hyb-mo/cs-hyb-mo_instructions-manual-and-user-manual_2698512008.webp)
8 |
P a g e
Curative Hybrid Very High Risk
Mattress
OPERATION
Pump Characteristics
Pump
P09 Pump
P09+ Pump
Power
AC220V / AC110V
AC220V / AC110V
Pressure Range
10
– 52 mmHg
10
– 28 mmHg
Internal Fuse
F0.5A 250V
F0.5A 250V
Plug Fuse
3A
3A
Frequency
50Hz / 60 Hz
50Hz / 60 Hz
Power
8W
8W
Noise
≤45dB
≤45dB
Alternating Time
1 complete cycle is 11 ½ minutes which is 2 x 5-minute periods
of alternative cell inflation with a change over time of 1 ½
minutes.
Pump Pressures
P09 Pump
P09+ Pump
Due to the low pressure and sensitivity of the pump, tolerances of 20% are accepted within
these models when conducting servicing and inspection protocol.
Mattress Characteristics
Pump
P09 Pump
P09+ Pump
Model
P09 pump & HH/MC Mattress P09+ pump & HH/MC Mattress
Layers
2
– Hybrid cells with U-core
2
– Hybrid cells with U-core
CPR
1
1
Max Loading Weight
180 Kg
180 Kg
Weight
(Kg)
Pressure
(mmHg)
Weight
(Kg)
Pressure
(mmHg)
30
10
110
26
40
12
120
28
50
14
130
28
60
16
140
28
70
18
150
28
80
20
160
28
90
22
170
28
100
24
180
28
Weight
(Kg)
Pressure
(mmHg)
Weight
(Kg)
Pressure
(mmHg)
30
10
110
26
40
12
120
28
50
14
130
30
60
16
140
32
70
18
150
34
80
20
160
40
90
22
170
46
100
24
180
52