Curative Hybrid Very High Risk 
Mattress 

OPERATION 

 
Working Principle 

The dynamic mattress system consists of a dynamic air pump and an alternating air 

mattress. The air mattress has two distinct airway chambers: 

 

1.  Group A Chambers 

– Air tubes 2,4,6 & 8. 

2.  Group B Chambers 

– Air tubes 1,3,5,7 & 9. 

 
A and B chambers alternate inflating and deflating by using the pump connected. This air 
alternating, mattress system enhances the blood circulation to treat and prevent pressure 
ulcers, when part of a holistic care package.  

 

Static Mode  

When the pump is adjusted to the static mode, air chambers A & B are both fully inflated until 
the maximum pump pressures are achieved. (Maximum pressures vary between different 
pump styles) Static mode is used for seating or patient transfer. 

 

 
Alternating Mode 

When started, both chambers A & B are initially inflated to the maximum pump pressure 
(Maximum pressures vary between different pump styles). Group A chambers deflate for 5 
minutes followed by both A & B inflating for 1.5 minutes. This is followed by Group B chambers 
deflating for 5 minutes. This is one complete cycle. 

 

 

Summary of Contents for CS-HYB-MO

Page 1: ...1 P a g e Curative Hybrid Very High Risk Mattress Instruction Guide User Manual Curative Hybrid Very High Risk Mattress Pump Mattress Model CS HYB MO Pump Model HYB 121 PUMP ...

Page 2: ...2 P a g e Curative Hybrid Very High Risk Mattress ...

Page 3: ... Principle OPERATION 5 Static Mode Alternating Mode Profiled Bed Clinical Considerations Installation Safety Instructions Pump Operation Mattress Pump Characteristics MAINTENANCE 9 Cleaning Maintenance Repair Storage Transportation Other Considerations Parts Accessories WARNINGS SYMBOLS 11 TROUBLE SHOOTING 12 COMPANY INFORMATION 13 ...

Page 4: ...ycle therefore providing direct pressure relief and patient stability throughout the whole dynamic process The multi stretch vapor permeable cover comprises of a high level coating of polyurethane which makes it easy to clean whilst complying with strict infection control procedures The rapid cell deflation system is unique in the sense that pressure relief happens much quicker than on a standard ...

Page 5: ...treat and prevent pressure ulcers when part of a holistic care package Static Mode When the pump is adjusted to the static mode air chambers A B are both fully inflated until the maximum pump pressures are achieved Maximum pressures vary between different pump styles Static mode is used for seating or patient transfer Alternating Mode When started both chambers A B are initially inflated to the ma...

Page 6: ...ce the pump either at the end of the bed on its hooks or on a flat surface 2 Ensure that there are no sharp or hard objects on the bed before unfolding the mattress laying out flat The air inlet should be located at the feet side of the bed 3 The tubes should be connected securely in place without any signs of leakage 4 Adjust the mode pressure by the user s condition weight to suit Safety Instruc...

Page 7: ...en pressure is normal extinguished when pressure is abnormal The light will continuously flash during the alternating cycle change over Air Leakage Low Pressure Alarm After working for 50 minutes and the pressure is lower than 15mmHg the pump will alarm indicating an initialisation failure During normal operation when the pressure is lower than 6mmHg for 6 minutes the pump will alarm indicating po...

Page 8: ...mp Due to the low pressure and sensitivity of the pump tolerances of 20 are accepted within these models when conducting servicing and inspection protocol Mattress Characteristics Pump P09 Pump P09 Pump Model P09 pump HH MC Mattress P09 pump HH MC Mattress Layers 2 Hybrid cells with U core 2 Hybrid cells with U core CPR 1 1 Max Loading Weight 180 Kg 180 Kg Weight Kg Pressure mmHg Weight Kg Pressur...

Page 9: ...ce repair work should only be carried out by an appointed engineer who have been trained by Care Shop and only original manufacturers parts should be used Failure to do so could result in a void warranty situation Storage Good ventilation will be required when storing the product it should always be kept dry away from direct sunlight Temperature should be maintained in the region of 22 C 3 C when ...

Page 10: ...fuse is disposed of in the appropriate manner The mattress should be away from sharp objects to avoid any scratches or punctures Put bed cover sheet or blanket over the mattress before use No plastic or rubber sheets should be used Ensure that all the tubes remain flat any twists kinks are avoided If any air leakage is found check all the tube connectors and air tubes carefully to see if any are l...

Page 11: ...st electric shock Class II electrical appliance Double insulated Dispose of electrical equipment correctly WEEE Do not use broken or damaged electrical plugs or cables as this may lead to electric shocks Ensure hands are dry before handling all electrical items No exposure to water splashing or it may cause electric shocks or fire Only authorised qualified personnel should dis assemble repair or r...

Page 12: ...d Check the tube connectors Repair the air tubes or call for replacement 3 Mattress is too hard or too soft The air pressure is not adjusting to required level Adjust the pump pressure according to the user weight until suitable and comfortable 4 No full inflation after long inflating Air hoses are loose or not connect properly Air pump has not enough air pressure Check all the air hoses and conne...

Page 13: ...ford Park Industrial Estate Warren Bruce Road Manchester M17 1LB Telephone 44 0 845 266 8744 Email enquiries careshop co uk Disclaimer Please be aware that all products should be disposed of in an environmentally friendly infection control manner in compliance with any local or national requirements ...

Reviews: