Curative Hybrid Very High Risk
Mattress
OPERATION
Pump Operation
Power:
Power On / Off
Soft / Firm:
Pressure adjustment. Each single push adjusts ±5KG. Longer push adjusts ±5KG
rapidly.
Static
: All tubes will inflate to maximum pump pressures (See page 8 for maximum pump
pressures). No alternating takes place & the digital display will show
“- - -“. The pump will
revert back to the previous alternating cycle after 30 minutes.
Pressure Monitor:
The light is illuminated when pressure is normal & extinguished when
pressure is abnormal. The light will continuously flash during the alternating cycle
change-over.
Air Leakage / Low Pressure Alarm
After working for 50 minutes and the pressure is lower than 15mmHg, the pump will
alarm indicating an initialisation failure.
During normal operation, when the pressure is lower than 6mmHg for 6 minutes, the
pump will alarm indicating possible air leakage from the mattress.
Pushing the mute key will silence the pump alarm.
If the alarm sounds consistently and all attempts to remedy have failed, seek engineer
assistance to rectify.
Power Off Alarm
The alarm will sound if power is lost to the pump.