4
V4-02-2011
TECHNICAL DATA
Measuring range
-80…50 °C pressure dew point resp. dew point in °Ctd
0…100 % RH
-30…70
°C
Type 0699.0401, FA 400
-20…50 °Ctd
≙
4…20 mA
Type 0699.0402, FA 400
-80…20 °Ctd
≙
4…20 mA
Type 0699.0403, FA 400
-60…30 °Ctd
≙
4…20 mA
Accuracy:
typical
±
2 °Ctd from -80…-40 °Ctd
±
1,5 °Ctd from -40…0 °Ctd
±
1 °Ctd from 0…20 °Ctd
Pressure range:
-1…50 bar standard, high pressure version up to 350 bar
Power supply:
24 VDC smoothed (16...30 VDC/60 mA)
Output:
4…20 mA 3-Draht-Technik
Protection class:
IP 65
EMV:
DIN EN 61326
Operating temperature:
0…50 °C
Storage temperature:
-40…80 °C
Burden for analogue output: < 500 Ohm
Screw-in thread:
G 1/2" stainless steel
Material housing:
Zinc alloy, PC, ABS
Sensor protection:
Sintered filter 50
μ
m stainless steel
Connection: M12,
5-pole
Response time t95:
< 30 seconds (dry)
< 10 seconds (humid)
Display:
4-digit
LCD
Alarm relay:
max. 60 V, 0.5 A, NO relay opens in case of alarm and power
failure, alarm value adjustable via keyboard, please see
page
9
Genauigkeit FA 410
0,0
0,5
1,0
1,5
2,0
2,5
3,0
-80
-60
-40
-20
0
20
Taupunkt [°C]
G
e
na
ui
g
k
e
it
[
°C
]
Accuracy FA 400
A
ccu
ra
cy
(°
C
)
Dew point (°C)
Summary of Contents for 0699.0401
Page 12: ...12 V4 02 2011 DECLARATION OF CONFORMITY...
Page 13: ...13 V4 02 2011 NOTES...
Page 14: ...14 V4 02 2011 NOTES...
Page 15: ...15 V4 02 2011 NOTES...