3.  Menu with all functions 

 
The operation is effected via 4 keys:         and           as well as                   ,  
 
in order to access the menu/to confirm selection and                    
 
The operation in detail: 
 
-45,0       currently measured value 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

press for the duration of 3 seconds 

AL - 1 

8.0 

alters 

At - 1 

nOr 

alters 

press shortly 

HY - 1 

2.0 

alters 

press shortly 

AL - 2 

8.0 

alters 

At - 2 

nOr 

alters 

press shortly 

HY - 2 

2.0 

alters 

press shortly 

rEv 

rEv 

With           alarm threshold 1 scaling up, 
 
with             scaling down  

nOr = Normal - means Alarm 1 
in case of "threshold exceeding" 

rEv = Reverse - means Alarm 1 
in case of "threshold undercut" 

With            Hysteresis 1 scaling up, 
 
with             scaling down 

     

 

     

 

With           alarm threshold 2 scaling up, 
 
with             scaling down  

     

 

nOr = Normal - means Alarm 2 
in case of "threshold exceeding" 

rEv = Reverse - means Alarm 2 
in case of "threshold undercut" 

With            Hysteresis 2 scaling up, 
 
with             scaling down 

     

 

Back 

 

 

Enter 

Enter 

Enter 

Enter 

Enter 

Enter 

Summary of Contents for 0500 0008

Page 1: ...V1_03_2016 1 Instruction Manual Digital Display DS 52 in wall housing Order no 0500 0008 for 0 4 20 mA sensors...

Page 2: ...gh standard products every year There are a few good reasons for doing so The cost performance ratio reliable quality at a fair price We have the ideal solutions for your measuring tasks based on our...

Page 3: ...the delivery capability or use of this device Initiation electrical connection When operating electrical instruments it is unavoidable that certain components are energised dangerously Work at electr...

Page 4: ...of CS Instruments sensors AC Power 230 VAC 50 Hz Attention Mains voltage NC und COM are closed in case of alarm exceeding voltage drop sensor break A bridge circuit is inserted ex factory For signal...

Page 5: ...ensor Terminal connector M12 7 8 9 101112131415 1 2 3 4 5 6 Sensor 230 VAC A1 2 A1 1 PE N L V GND I I I SIG 1 2 3 4 5 FA 515 Braun brown Blau blue CS Sensor Terminal connector M12 A bridge circuit is...

Page 6: ...or input is configurated on our premises according to the customers specification in the order Sensors with 0 4 20 mA current signal are connected as follows V max 24 VDC 100 mA A bridge circuit is in...

Page 7: ...9 101112131415 1 2 3 4 5 6 Sensor 230 VAC A1 2 A1 1 PE N L V GND I I I SIG 0 1 2 5 4 3 Netzanschlussleitung Supply cable gr n gelb green yellow blau blue braun brown grau grey schwarz black Lampe lamp...

Page 8: ...2 nOr alters press shortly HY 2 2 0 alters press shortly rEv rEv With alarm threshold 1 scaling up with scaling down nOr Normal means Alarm 1 in case of threshold exceeding rEv Reverse means Alarm 1...

Page 9: ...with and whether you want to connect a sensor with 0 20 mA signal or with 4 20 mA signal 4 mA 0 20 mA 500 With and you can set the lower scaling value 4 mA resp 0 mA of your sensor With and you set th...

Page 10: ...instrument has been opened as far as this has not been mentioned in the instruction manual for maintenance purposes or if the serial number in the instrument has been changed damaged or removed The wa...

Page 11: ...or sensors with 0 20 mA or 0 4 20 mA signal Connectable in 2 3 or 4 wire technologie possible with 24 VDC supply Burden analogue input 100 Ohm Accuracy analogue input max 20 A typical 10 A Sensor supp...

Page 12: ...ogue sensor input for sensors with 4 20 mA or 0 20 mA signals 2 alarm relays change over contact 250 VAC max 3 A alarm values freely adjustable 0500 0008 Options Option for DS 52 LED display Supply 24...

Page 13: ...V1_03_2016 13 Notes...

Page 14: ...55 Harrislee Phone 49 0 461 700 2025 Fax 49 0 461 700 2026 E Mail info cs instruments com www cs instruments com Sales office South Germany CS Instruments GmbH Zindelsteiner Stra e 15 78052 VS Tannhei...

Reviews: