22
2.8.3 Calibration after bump fail Recommended Gas Levels
shows the recommended minimum and maximum concentration of gas suitable for a Gas-
Pro when configured to undertake a bump fail calibration, when the unit fails either a speedy or a
smart bump.
Table 3
Gas
Gas Name
Recommended
Minimum
Concentration
Recommended
Concentration
Recommended
Maximum
Concentration
O
2
Oxygen
≥ 18% VOL
18% VOL
≤ 22% VOL
CO
Carbon Monoxide
≥ 100ppm
250ppm (100ppm)
≤ 400ppm
H
2
S
Hydrogen Sulphide
≥ 15ppm
25ppm (15ppm)
≤ 85ppm
CO
2
Carbon Dioxide
≥ 0.5% VOL
2% VOL (1%)
≤ 2% VOL
SO
2
Sulphur Dioxide
≥ 10ppm
10ppm
≤ 16ppm
CH
4
Methane
[ATEX] 2.2% VOL
[ATEX] 2.2 % VOL
[ATEX] ≤ 2.2% VOL
[UL] 2.5% VOL
[UL] 2.5% VOL
[UL] ≤ 2.5% VOL
C
4
H
10
Butane
[ATEX] 0.7% VOL
[ATEX] 0.7% VOL
[ATEX] ≤ 0.7% VOL
[UL] 0.6% VOL
[UL] 0.6% VOL
[UL] ≤ 0.6% VOL
C
3
H
8
Propane
[ATEX] 0.85% VOL
[ATEX] 0.85% VOL
[ATEX] ≤ 0.85% VOL
[UL] 1.05% VOL
[UL] 1.05% VOL
[UL] ≤ 1.05% VOL
C
5
H
12
Pentane
[ATEX] 0.55% VOL
[ATEX] 0.55% VOL
[ATEX] ≤ 0.55% VOL
[UL] 0.75% VOL
[UL] 0.75% VOL
[UL] ≤ 0.75% VOL
C
2
H
2
Acetylene
[ATEX] 1.15% VOL
[ATEX] 1.15% VOL
[ATEX] ≤ 1.15% VOL
[UL] 1.2% VOL
[UL] 1.2% VOL
[UL] ≤ 1.2% VOL
C
2
H
4
Ethylene
[ATEX] 1.15% VOL
[ATEX] 1.15% VOL
[ATEX] ≤ 1.15% VOL
[UL] 1.35% VOL
[UL] 1.35% VOL
[UL] ≤ 1.35% VOL
NH
3
Ammonia
≥ 50ppm
50ppm
≤ 85ppm
N0
2
Nitrogen Dioxide
≥ 10ppm
10ppm
≤ 17ppm
NO
Nitric Oxide
≥ 50ppm
50ppm
≤ 85ppm
Contents
Exit
Prologue
Set-up
Operation
Service and
maintenance
Specification
Warranty
Unpacking
Contacts
Troubleshooting