2.6 Menu Settings
All program changes must start with the operation of the Menu/Execute switch and must also end with the
operation of the Menu/Execute switch verified by the “Save Settings?” Menu. If this sequence is not followed,
none of the changes will take effect. If programming is initiated and no operator action takes place for
approximately 30 seconds (before the final press of the Menu/Execute switch) the display will revert to its
previous status and you will need to start over.
Alarm indications appear on the LEDs (See figure 2.3). An alarm condition will occur if any local oscillator
phase lock loop (PLL) comes out of lock. The Remote LED will light when you select the Remote mode.
Figure 2.5 Front Panel Menu Display & Sequence
OTHER SETTINGS
EXT REF = INT
SAVE SETTINGS Y N
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
SCROLL
OTHER SETTINGS
REMOTE = ON
OTHER SETTINGS
INTERFACE = RS232
RS485 ADDR = 00
R
R
R
R
PUSHING MENU/EXECUTE SEQUENCE
ON POWER UP
SCROLL <>
SCROLL
IP Address
Model Number
Firmware Rev.
Normal Display
Menu 1 Select Settings
Menu 2 Set Frequency
Menu 3 Set Input Level
Menu 4 Select Gain
Menu 5 Select Mute
Menu 6 Reference Insert
Save? When “R” is selected
from any above menu or
at the end.
NORMAL DISPLAY
WHICH SETTINGS?
#1 #2 #3 #4 OTHER
CH1 UPCONVERTER
FREQ= 950.000
R
1674.125 1926.250
1926.250 1250.125
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
SCROLL
SCROLL <>
SCROLL
CH1 UPCONVERTER
INPUT LVL = -20
CH1 UPCONVERTER
GAIN = +20.0
CH1 UPCONVERTER
MUTE = OFF
CH1 UPCONVERTER
REF INSERT = OFF
R
R
R
R
R
SAVE SETTINGS Y N
192.168.123.2
REPEAT ABOVE SEQUENCE
FOR ALL INDIVIDUAL CHANNELS
2415-402
5.00
2415-x02 Manual, Rev. E