Sensor:
State of the art E_MOS
®
technology.
Gases:
Detects refrigerants including those with
hydrocarbon components (CFC, CH2F2,
HCFC, HFC, HFO, NH3, HC. Hydrogen
trace gas in 95% N2, 5% H2 mixture A2L
refrigerants).
CAUTION: This leak detector is not
intrinsically safe. Use in well ventilated areas
.
Sensitivity:
Better than a 0.1 oz / year (3 g / annually) for
all common refrigerants, including R-410a,
R-404a and R-407a.
Audible Alarms:
A bar-graph LED display shows the relative
leak size in each of 7 possible ranges. Each
range is indicated by a numbered LED. Each
range is automatically selected by the instru-
ment starting with highest sensitivity and
automatically switching to lower sensitivities
as the leak source is approached.
Operating
Temperature:
-18 to 60˚C (0 to 140˚F)
Power:
Three size "C" alkaline batteries (NEDA/
ANSI 14A).
Battery Life:
50 hours minimum at 77˚F (25˚C). Battery
life depends on cell chemistry and ambient
conditions. Alkaline batteries provide longest
life. Cold ambient conditions decrease bat-
tery life. The
Leak-Seeker
®
II
turns itself off
after 10 minutes of inactivity—
START
button
not pressed—to further prolong battery life.
Dimensions:
Instrument (L x W x H): 10" x 2.3" x 2"
254mm x 59mm x 51mm
Probe:
15" (380mm) long, .25" (6.4mm)diameter.
Weight:
1 lb 2 oz (500 g)
Certifications:
Complies with European Standard
EN14624:2012
Leak-Seeker
®
II
Specifications
3
GENERAL INFORMATION