Installation
CPI Canada Inc.
2.8.4 Inspecting DIP switch SW8 setting for DSS Board #903132-xx
DIP switch SW8 on the dual-speed starter must be set correctly. Failure to do this may result
in damage to the X-ray tube or generator.
SW8 is set at the factory. Adjustment of this switch is not required. However it is recommended to
inspect the switch to verify that it is set properly before proceeding further. The settings are as follows:
SW8
- switches
1,2,3,4,6,7 - OFF
SW8
- switches
5,8 - ON
NOTE: ONLY CHECK THE SWITCH SETTINGS OF DIP SWITCH SW8. MAKING ADJUSTMENTS
DIFFERENT FROM THE ABOVE SETTING WILL RESULT IN IMPROPER ANODE RPM
WHICH CAN CAUSE OVERHEATING AND/OR DAMAGE TO THE X-RAY TUBE.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 2-26
CMP 200
X-Ray Generator Service Manual Ch # 901476-02 Rev. P