Connecting the X-ray Tube Stator Cable for Dual-speed Starter
The connector is designed for connecting the X-ray stator cable for the dual-speed starter is located on
the back of the DSS board mounting plate as shown in Figure 2-6.
Figure 2-6: the connector of the X-ray stator cable for the dual-speed starter
1.
Route the X-ray tube stator cable towards the X-ray stator cable connector on the back of the
DSS board mounting plate. Refer to figure 2-6.
THE SHIELD FOR THE STATOR CABLE MUST BE PROPERLY GROUNDED AT BOTH THE
TUBE END AND THE GENERATOR END OF THE CABLE.
2.
Connect the shield ground of the stator cable to the ground connector on the mounting plate as
shown in Figure 2-6.
3.
Connect the X-ray tube stator cable to the X-ray stator cable connector as shown in Figure 2-6.