Please make sure the Power Supply is OFF before connecting/disconnecting the COVVI Hand to and from the
prosthesis as this will avoid any unpredictable spikes in the electric current from the batteries. Turn the power
on once the hand is properly connected to the prosthesis.
Please see the image below for instructions on how to connect the COVVI Hand to your patient’s socket.
6.1 Connecting/Disconnecting the COVVI Hand
6.0 Fitting the COVVI Hand
If the ball bearings are locked out radially, the hand will not plug in.
Remove protective cap
Ensure no debris is inside
the EQD core before
connecting the hand to
the socket.