Connecting the Nurse Call Cable
Puritan Bennett 540™ Ventilator User’s Manual
4-19
•
The ventilator signals an alarm using a normally open (NO) or a normally closed (NC) signal.
•
A remote alarm is activated when an alarm condition occurs, unless either of the following is true:
•
The audio paused function is active.
•
The ventilator power switch is OFF.
•
The remote alarm port is an 8-pin female connector; allowable current is 100mA at 24VDC(max).