Performance
Puritan Bennett 540™ Ventilator User’s Manual
A-3
A.4
Performance
A.4.1 Specifications
A.4.2 Test Results
Table A-5.
Power Indicators
Ventilation ON/OFF
AC power
DC power
INTERNAL BAT
•
Blue in standby mode
•
Not lit if ventilation is in
progress.
Green
Green
•
Flashing if the battery
charge is in progress.
•
Continuously lit if the
ventilator is powered by
the internal battery.
Table A-6.
Alarm Indicators
High Priority
Medium Priority
Red flashing LED
Yellow flashing LED
Table A-7.
Audio Alarms
Audio Paused
Alarm Volume
60 s ± 1 s
65 to 85 dBA ± 10% at 1 meter
Table A-8.
Performance Parameter Specifications and Tolerances
Settings Range
Tolerances
Volume
50 to 2000 ml
± (10 ml + 10%)
Pressure
5 to 55 cmH
2
O
± (1 cmH
2
O ± 10%)
Time
0.3 to 6.0 s
± 10%
Rate
1 to 60 bpm
± 1 bpm
Inspiratory Sensitivity
1P to 5
N/A
Exhalation Sensitivity
5 to 95%
N/A
Table A-9.
Volume Performance Test Results
R-Rate
(bpm)
5
10
15
20
30
40
50
60
Insp Time
(s)
4.0
2.0
1.3
1.0
0.7
0.5
0.4
0.3
C50 R5
VTI min. (ml)
220
110
80
60
60
NA
NA
NA
VTI max. (ml)
2000
2000
1750
1200
700
NA
NA
NA
C20 R20
VTI min. (ml)
NA
110
80
60
60
NA
NA
NA
VTI max. (ml)
NA
1000
900
700
400
NA
NA
NA
C20 R20
VTI min. (ml)
NA
NA
NA
VTI max. (ml)
NA
NA
NA
Summary of Contents for Puritan Bennett 540
Page 1: ...Puritan Bennett 540 Ventilator User s Manual...
Page 6: ...This page intentionally blank...
Page 8: ...This page intentionally blank...
Page 10: ...This page intentionally blank...
Page 94: ...This page intentionally blank...
Page 118: ...Specifications A 18 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Page 123: ...Puritan Bennett 540 Ventilator User s Manual C 3 Figure C 2 Dual Bag...
Page 134: ...Glossary F 6 Puritan Bennett 540 Ventilator User s Manual This page intentionally blank...
Page 140: ......