Specifications
88
Table 15: Electromagnetic Emissions
The N-560 is intended for use in the electromagnetic environment specified below.
The customer or the user of the N-560 should assure that it is used in such an
environment
Emission Test
Compliance
Electromagnetic Environment —
Guidance
RF emissions
CISPR 11
Group 1
The N-560 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The N-560 is suitable for use in
establishments, including diagnostic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Table 16: Electromagnetic Immunity
The N-560 is intended for use in the electromagnetic environment specified below.
The customer or user of the N-560 should assure that it is used in such an
environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment —
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Complies
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV power
supply lines
±1 kV for
input/output
lines
Complies
Mains power should be
that of a typical
commercial or hospital
environment.
Summary of Contents for Nellcor OxiMax N-560
Page 1: ...Service Manual OxiMaxTM Technology Nellcor OxiMax N 560 Pulse Oximeter TM...
Page 10: ...List of Tables N 560 Service Manual viii...
Page 38: ...Performance Verification 28...
Page 80: ...Disassembly Guide 70 Figure 41 Equipotential Ground Terminal Removed...
Page 83: ...Spare Parts N 560 Service Manual 73 Figure 42 Exploded View...
Page 84: ...Spare Parts 74...
Page 88: ...Packing for Shipment 78...
Page 104: ...Specifications 94...
Page 114: ...Data Port Interface Protocol 104...
Page 128: ...Technical Discussion 118...
Page 130: ...Technical Discussion 120...
Page 132: ...Technical Discussion 122...
Page 134: ...Technical Discussion 124...
Page 136: ...Technical Discussion 126...
Page 138: ...Technical Discussion 128...
Page 139: ...Technical Discussion N 560 Service Manual 129 Figure 55 Main PCB Parts Locator Diagram...
Page 140: ...Technical Discussion 130...
Page 142: ...Technical Discussion 132...
Page 144: ...Technical Discussion 134...
Page 146: ...Technical Discussion 136...
Page 147: ...Technical Discussion N 560 Service Manual 137 Figure 59 Front Panel Parts Locator Diagram...
Page 148: ...Technical Discussion 138...
Page 149: ...Technical Discussion N 560 Service Manual 139 Figure 60 Power Supply Schematic Diagram...
Page 150: ...Technical Discussion 140...
Page 151: ...Technical Discussion N 560 Service Manual 141 Figure 61 Power Supply Parts Locator Diagram...
Page 152: ...Technical Discussion 142...
Page 156: ......