Coulometrix Metiri COUMET60100SB User Manual Download Page 2

TECHNICAL SPECIFICATION

Model Number

COUMET60100HB

COUMET60100SB

Maximum Operating Values

Voltage Operation Range

+5.5Vdc to +60.0Vdc

+6.5Vdc to +60.0Vdc

...with Aux Power Supply

0.0Vdc to +60.0Vdc

0.0Vdc to +60.0Vdc

Maximum DC Current

+/-70Adc sustained

+/-70Adc sustained

Maximum Peak Current

+/-170A for less than 3 seconds

+/-170A for less than 3 seconds

Input Current from Source or Load

28mA +/-10%

8mA to 149mA*

...with Aux Power Supply

<1mA

<1mA

Measurement Capabilities

Voltage Detection Resolution

10mV

10mV

Current Detection Resolution

100mA

100mA

Current Measurement Accuracy

+/-1% from 2A ~ 170A**

+/-1% from 2A ~ 170A**

Sample Resolution

8Hz (every 125mSec)

8Hz (every 125mSec)

Clock Resolution

+/-2%

+/-0.5%

Max Energy Recordable Value

+/-2,147,483,648 Joules (watt 

seconds)

+/-2,147,483,648 Joules (watt 

seconds)

Max Clock Recordable Value

65,536 days

65,536 days

Recording Capability

MicroSD Card Size Capability

n/a

FAT32, up to 128GB

SELECT or ENTER Screen Timeout

10 seconds

10 seconds

* Device current depends on source voltage.  Higher the voltage, the lower the current.
** Current accuracy depends on proper calibration.

PRODUCT OPERATION
Connections:

  The Metiri™ Smart Energy Meter prod

uct line is an inline voltage, current, and energy measurement 

device with all of the features and tools a solar enthusiast, RC hobbyist, or Scientific community would love.

The Metiri™ Smart Energy meter is in series with a pair of pure copper 10 AWG wires, one 

RED

 and one 

BLACK

 in 

color.  The 

RED

 wire is the positive or power wire and the 

BLACK

 wire is the return or ground wire.  Metiri™ measures 

voltage between these wires and uses a low side resistive shunt for monitoring current.  The two exposed wires closer 

to the USB connector should be connected to a power source or battery, and the two exposed wires farthest from the 
USB connector should be connected to a power load or battery.  

CAUTION

:

  Do not place a negative voltage or reverse the voltage polarity of the wires or damage to the Metiri™, and

your system, might occur.  

WARNING

:

  Always install and in line fuse rated for no more than 200A in series with the Metiri™ Smart Power Meter

Device Power:

  The Metiri™ Smart Energy Meter will get the power it needs to run from the wires it is measuring and 

requires a minnimum voltage accross those wires to function.  If a voltage is desired to be measured that is lower than
the minimum voltage, auxilliary power will be needed.  The Metiri™ products provide an auxillary power input using a 

readily available mini USB connector.  Simply plug in a mini USB cable into Metiri™, and the other end into any USB 
power source, and Metiri™ will shut down its internal regulator, running off of the USB power only.

Device Measurement:

  When opperating Metiri™ from the Auxilary USB, the Source or Load cables should not be left

unconnected (otherwise known as high impedance) or leakage will cause incorrect measurements to occur.
In order to take accurate current measurements, it is important to understand that the minimum current resolution is a
percentage of the maximum current the unit is designed to read.  The higher the current capability desired, the lower 

the sense resistance needed to keep from overheating and thus less signal to noise.

Safe Operating Current:

  The Metiri™ products are fully bi-directional and will record current flowing in either 

direction.  This makes the Metiri™ products perfect for monitoring energy flowing in and out of a system battery bank. 
The Metiri™ products attempt to interfear as little as possible with current moving through it.  Therefore the user is 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for Metiri COUMET60100SB

Page 1: ...ted with FAT32 Metiri S will automatically detect card create a CVS file and start writing measured data 6 Source Wires Connect to power source or battery up to 60Vdc Current flowing from Source to Lo...

Page 2: ...power source or battery and the two exposed wires farthest from the USB connector should be connected to a power load or battery CAUTION Do not place a negative voltage or reverse the voltage polarit...

Page 3: ...urn to after 10 seconds of inactivity in any screen except when scrolling through the Smart Energy Graph screen All measured values presented are in real time and updated at the same rate as the sampl...

Page 4: ...currently programmed for the Home screen Touching SELECT will change the units to be displayed and will save this setting to the User Memory Touching ENTER after making a selection will return to the...

Page 5: ...shown below If the negative current calibration is desired touch ENTER once more and instructions for Negative Current Voltage A Calibration will show STEP 3 Connect 12 volts to the SOURCE wires make...

Page 6: ...d using 8 3 filenames The Metiri S names files according to the current day proceeded by the letter M followed by two letters starting at AA If that filename is already taken then the prefix letters w...

Page 7: ...higher than the stated values will decrease the Safe Operating Current levels and time 4 Never touch the wires when operating the unit They may be hot and can cause burns to exposed skin or material...