Metiri™ Scientific COUMET60100SB

Metiri™ Hobbyist COUMET60100HB

Smart Energy Meter

Users Manual for Firmware ver: 0.8.8

QUICK PRODUCT EXPLAINATION

1.

LCD

:  Back-lit primary display.  

2.

Select Button

:  Touch sensitive button with LED touch indicator.  Press to scroll through data options or Smart

Graph.

3.

Enter Button

:  Touch sensitive button with LED touch indicator.  Press to return to the main screen.  If on the 

main screen, will scroll through set-up options.

4.

Aux. Power

:  Female socket for device auxilliary power, 5.0Vdc +/-10%.  Compatable with USB mini-B Plug.

5.

MicroSD Slot (Scientific Model Only)

:  Card slot for recording of measured values.  Can accept standard 

microSD flash memory formated with FAT32.  Metiri™-S will automatically detect card, create a CVS file, and 
start writing measured data.

6.

Source Wires

:  Connect to power source or battery up to +60Vdc.  Current flowing from Source to Load is 

considered positive current.  Forward continuous current should never exceed 70 amperes.

7.

Load Wires

:  Connect to power load or battery up to +60Vdc.   Current flowing from Load to Source is 

considered negative current.  Negative continuous current should never exceed 70 amperes.

8.

In Line Fuse (Sold Separately)

:  The Metiri™ products do not provide protection against too much current 

flowing through it.  A fuse installed in series with the positive Source Wire is required.  Current flowing through 

the Metiri™ product should never exceed the current specified in the Safe Operating Area graph.

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for Metiri COUMET60100SB

Page 1: ...ted with FAT32 Metiri S will automatically detect card create a CVS file and start writing measured data 6 Source Wires Connect to power source or battery up to 60Vdc Current flowing from Source to Lo...

Page 2: ...power source or battery and the two exposed wires farthest from the USB connector should be connected to a power load or battery CAUTION Do not place a negative voltage or reverse the voltage polarit...

Page 3: ...urn to after 10 seconds of inactivity in any screen except when scrolling through the Smart Energy Graph screen All measured values presented are in real time and updated at the same rate as the sampl...

Page 4: ...currently programmed for the Home screen Touching SELECT will change the units to be displayed and will save this setting to the User Memory Touching ENTER after making a selection will return to the...

Page 5: ...shown below If the negative current calibration is desired touch ENTER once more and instructions for Negative Current Voltage A Calibration will show STEP 3 Connect 12 volts to the SOURCE wires make...

Page 6: ...d using 8 3 filenames The Metiri S names files according to the current day proceeded by the letter M followed by two letters starting at AA If that filename is already taken then the prefix letters w...

Page 7: ...higher than the stated values will decrease the Safe Operating Current levels and time 4 Never touch the wires when operating the unit They may be hot and can cause burns to exposed skin or material...

Reviews: