Maintenance Inspection
■
Daily Inspection
■
Regular Inspection
In order to maintain the accuracy of the Gas Detector, request a regular inspection from your New-
Cosmos Electric representative.
All XP-3000 series gas detectors which are calibrated have the same algorithm to calculate the
measurement data to the concentration.
■
Main Replacement Parts
6. Maintenance Inspection
Check item
Description
Gas sampling hose
Make sure that the gas sampling hose has no bends or damage and
that it is securely connected.
Drain filter or
Cooling drain filter
If the filter element in the drain filter is dirty or wet or water remains
in the drain filter, clean the inside of the drain filter (and 1-m-long gas
sampling hose) and replace the filter element with a new one (see
page 18).
Remaining battery level
If the remaining battery level is low, replace the batteries with new
ones (see page 19).
Airtightness
Press the power switch ON to display the gas concentration
screen, cover the gas sampling pipe tip with a finger and check
the error message "Err-P"(blockage malfunction) is displayed.
As "Err-P" is displayed, it is a normal state and operable by
pressing the power switch ON again. if "Err-P" is not displayed
the damage of the gas sampling pipe or the airtight defect of the
sensor packing or pump diaphragm may be considered.
Contact your New-Cosmos Electric representative for repairs or
replacements.
Product
Model
Remarks
Filter Element
FE-2 (10 filters)
For XP-3180
Filter Element
FE-10 (10 filters)
For XP-3180E
Oxygen sensor
OS3M-SS
For XP-3180
Oxygen sensor
OS3ME-SS
For XP-3180E