background image

CA 8010 Instruction Manual 

 
 

 

 

                          

         - 9 - 

 

Calibration 

Once you have made the settings you require, and when no key is pressed for a short 
time the display shows „SH 99“. After the heating-up phase of the sensor the display 
shows „BLOW“. You are now ready to begin with gassing the sensor. 
 

1.

 

Run: 

In the first step you gasing the sensor with a calibration solution of 0,20‰.  
Please confirm the measuring result which is indicated by pressing the OK 
button.  

 

2.

 

Run: 

Start the second run by pressing the OK button. The display shows again  
„SH 99“, the sensor heats up and the display shows again “BLOW”. Now you 
start with the second gassing by using a calibration solution of 1,00‰. Please 
also confirm the second measuring result and finalise the calibration 
procedure by pressing the OK button. 
 

After the completetion of this works the CA 8010 is ready for operation for the next 
6 month. 

Order information – for the CA 8010

 

* VAT to be added to all prices plus shipping costs, if necessary. Subject to change. 

Product 

Number 

Item no. 

Price* 

Breathalyzer CA 8010 black 

Shipment in a aluminium case, with accessories comprising:  
25  mouthpieces, batteries, belt pouch, instruction manual 

1. piece 

8010 

 

 

 

 

Mouth pieces with rebreathing inhibitor 

1 VE = 25 St. einzeln verpackt

 

1. piece 

0023 

 

 

 

 

 

Belt pouch 

For the convenient storage of the device and mouth pieces  

1. piece 

7202 

Transport- and storage case 

1. piece 

 

USB jack 

For  the connection of the Alco Conect 8010 to a PC or Laptop 

1. piece 

 

 

 

 

 

Gas simulator exclusiv 

Complete with thermostat for constant temp. 34°C, air pump for 
calibration, flow meter with control and automatic air-flow stop. 

1 piece 

0403 

Calibration-alcohol-solution  

for Cosmos Gas simulators  500 ml with  0,20 ‰ + 1,00 ‰  

1 bottle 

0404 

Summary of Contents for CA 8010

Page 1: ...2 WORD A4 E 10 10 C A 8010 GA Kal GA Alle Sprachen...

Page 2: ...ion the AlcoConnect 8010 7 Switch on the device 8 Calibration mode 8 Settings 8 Country code 8 Measuring unit 8 Alert level 8 Calibration 9 Order information for the CA 8010 9 Troubleshooting 10 For y...

Page 3: ...measurements 3 Measuring result in 4 Simple and fast calibration can also be performed by the client himself To do that Cosmos provides 1 mobile devices Gas simulator For further information please a...

Page 4: ...lead to criminal culpability In Germany criminal prosecution and a fine can be expected if neurological impairment is 0 30 or above A test with the CA 8010 will not be accepted by the police as eviden...

Page 5: ...s maintained in the case of the rechargeable battery runs low Loading the rechargeable battery Please connect the CA 8010 to the supplied USB cable and a free USB port on your PC Notebook or car adapt...

Page 6: ...tes before a test 1 Switch on the CA 8010 by press briefly the power button 2 By pressing the OK button for more than one second the display shows PH Now the sensor is pre hated 3 Please place the mou...

Page 7: ...to the time 199 measuring results will be stored if the memory is full the oldest result is automatically deleted Time and date adjustment Please press the lateral down key for 3 seconds The second in...

Page 8: ...00 with the number 20 by pressing the key and confirm with OK Now the third and fourth 00 flashes Replace these two 00 with the number 10 by pressing the key Please do not press the OK button or any o...

Page 9: ...essing the OK button After the completetion of this works the CA 8010 is ready for operation for the next 6 month Order information for the CA 8010 VAT to be added to all prices plus shipping costs if...

Page 10: ...ries do not throw into fire do not charge batteries Risk of explosion do not use force to open Risk of explosion Only dispose of as toxic waste Comply with local regulations governing disposal For you...

Page 11: ...l sensor fuel cell Precision 0 05 at 0 50 Measurement range 0 10 to 4 00 0 05 to 2 00 mg L Temperature range 5 C bis 50 C Shutdown automatic after 30 seconds The CA 8010 should be recalibrated every 6...

Page 12: ...de without authorisation from us Applicable EC guidelines Emissions Radiated emission EN55022 Interference immunity Electrostatic discharge EN61326 1 Electromagnetic field EN61326 1 Magnetic field pow...

Reviews: