パワーサプライや電気泳動槽の汚れを落とす際には、中性洗剤を含ませたよく絞った
布などで拭いてください。有機溶媒を使用すると変形、ヒビ割れの原因となります。
When cleaning the Power supply and Gel electrophoresis tank, wipe them using a 
soft cloth, dampened with mild detergent. Usage of organic solvent may cause 
deformation or create cracks.

バッファーの濃度が濃すぎたり、バッファーの液温が適正温度(40 度以下)を越えて、
負荷がかかり過電流となった場合、クロスオーバー機能が働きます。

*クロスオーバー機能:サンプルの変性を抑えるため自動的に定電流制御に切り替わります。急激な変化は
検知され、E1 エラー(過電流エラー)になります。

Crossover function activates when in cases as buffer concentration becomes too high, 
or overcurrent event occurs due to buffer temperature rose over proper level (over 40 
degrees Celsius [℃]/104 degrees Fahrenheit [℉]).

* Crossover function: Automated function to suppress sample aggregation. Sudden current change will be 
detected and E1 error (overcurrent error) will show in the display.

電気泳動槽に長期間バッファーを入れたまま放置しないでください。電極の劣化のおそ
れがあります。
Do not leave buffer in Gel electrophoresis tank for prolonged period, as degradation of 
electrode could result.

電気泳動槽およびゲルキャスティングトレイは紫外線透過性の素材を用いていますが、
長時間の照射により変色、紫外線透過率が低下します。また、紫外線から眼、皮膚を
保護するためのバイザーやゴーグルなどを使用してください。
Gel electrophoresis tank and Gel casting tray both have UV (Ultraviolet) light 
transparency, though prolonged exposure to UV light may cause discoloration or 
decrease of UV light transparency level. Always wear a visor or approved goggles to 
help vision and to protect your eyes and skin from UV light. 

核酸染色液は人体に害を及ぼす危険性がありますので、使用する際は必ずグローブを着
用し、皮膚に触れないようにしてください。
Always wear suitable protective gloves and avoid skin contact with nucleic acid stain, as 
it may be hazardous to human body.

電気泳動槽は丸洗い可能ですが、洗浄後はしっかりと乾燥させてください。濡れたまま
使用すると、感電・火災・故障の原因となります。
Gel electrophoresis tank is washable, though dry well after the wash, as electrical shock, 
a fire or equipment damage could result when using it wet.

オートクレーブを使用すると変形、故障の原因になりますので使用しないでください。
Never autoclave this product, as deformation and equipment damage could result.

使用上の注意 OPERATING PRECAUTIONS !

5

i-MyRunⅡ核酸電気泳動システム取扱説明書̲B5.indd   5

i-MyRunⅡ核酸電気泳動システム取扱説明書̲B5.indd   5

2014/10/09   14:49:56

2014/10/09   14:49:56

Summary of Contents for i-MyRun II

Page 1: ...i MyRun II...

Page 2: ......

Page 3: ...N SIGN WARNING SIGN Electricity Indicates instructions to follow Important user information Indicates a potentially hazardous situation which if not avoided will result in death or serious injury Indi...

Page 4: ...l shock Do not use this product if outer damage cuased by strong shocks such as from being dropped as a re or electrical shock could result Do not operate this product nor unplug Power cable from the...

Page 5: ...bu er in it separate the Power supply rst before moving unless hold the Power supply and the tank rmly with both hands as electrode connectors will be stressed equipment damage or dropping of product...

Page 6: ...eave bu er in Gel electrophoresis tank for prolonged period as degradation of electrode could result Gel electrophoresis tank and Gel casting tray both have UV Ultraviolet light transparency though pr...

Page 7: ...Features Power supply Gel electrophoresis tank Gel electrophoresis tank cover Gel casting tray Combs Gel casting stand Power cable 1 1 Standard components 1 1 each 1 1 each 1 1 each 1 1 each 6 6 each...

Page 8: ...rom Gel electrophoresis tank Power supply will detach from the Gel electrophoresis tank by rotating towards it How to attach Power supply and Gel electrophoresis tank When attaching Power supply slide...

Page 9: ...ns 0 99 0 Timer settings from 0 99 minutes Gel electrophoresis will end when timer reaches 0 zero When timer is set to 0 zero will be displayed for continuous run mode 3 Voltage setting button 50 75 1...

Page 10: ...or 0 5 TAE bu er is recommended For preparation of 4 mm thickness agarose gel amount of 70 80 mL gel will be required For 4 mm thickness agarose gel 9 L 26 well comb or 18 L 13 well comb of sample ca...

Page 11: ...tore agarose gel you have prepared be sure to cover the gel with bu er after gently drawing out Comb s Wrap the gel in plastic wrap and store it under low temperature Remove Gel casting tray with agar...

Page 12: ...electrophoresis Please double check that RUN button lights up and air bubbles forming from platinum wires at both ends of the tank Performing gel electrophoresis for prolonged time under condition of...

Page 13: ...t taking out the gel from the tank if staining dye has been added into the agarose gel when preparing it Always wear a visor or approved goggles to help vision and to protect your eyes and skin from U...

Page 14: ...W 65mm H 240 mm L Gel electrophoresis tank UV light transparent resin Gel casting tray UV light transparent resin Gel electrophoresis tank cover UV light un transparent resin Not enough current in th...

Page 15: ...d strains tilts Make sure that agarose gel does not contain any air bubbles nor undissolved gel blocks Make sure that agarose gel is not too thin For preparation of 4 mm thickness agarose gel amount o...

Page 16: ...If observing the result through gel electrophoresis tank or tray try observing only the gel If the gel after gel electrophoresis has been left out for prolonged period nucleic acid bands may di use an...

Page 17: ...memo 16...

Page 18: ...i MyRun II e mail FAX 1 i MyRun II IMR2 001 FAX TEL e mail 17...

Page 19: ...1 2 3 4 18...

Page 20: ...0016 2 2 20 TEL 03 5632 9610 FAX 03 5632 9619 TEL 03 5632 9620 URL http www cosmobio co jp TOYO EKIMAE BLDG 2 20 TOYO 2 CHOME KOTO KU TOKYO 135 0016 JAPAN TEL 81 3 5632 9617 FAX 81 3 5632 9618 email...

Reviews: