3
Operating the Corning
LSE
Vortex Mixer
All operating controls are located on the front panel. The power switch, located on the left side of the
panel, has three positions - TOUCH, OFF and ON. When set to the TOUCH position, the mixer is
activated by depressing the mixing head. When pressure is released from the mixing head, operation will
stop. When set to the ON position, the mixer will operate continuously until the switch is moved to either
the OFF or TOUCH position.
The speed control, located on the right side of the panel, is continuously adjustable. Oscillating speed
increases at a nonuniform rate to a maximum of 3,400 rpm (2850 for 230V units) as the knob is turned
clockwise. Before turning on the mixer, set the speed control to the minimum.
To begin operation, move the power switch to either the TOUCH or ON position and turn the speed
control to the desired setting. If the power switch is in the ON position, the shaking head will begin
moving immediately. If the power switch is in the TOUCH position, operation will begin when an object is
placed on the shaking head and slight pressure is applied to depress the head. Shaking will cease when
the object is removed and pressure released.
Preventative Maintenance
Warning: Disconnect Corning
LSE
Vortex Mixer from power source before
attempting any cleaning or maintenance procedures.
After each use, wipe down the mixer with a soft cloth. Do not immerse the unit or pour liquids over it as
electrical shock may occur. Clean up any spills immediately taking the proper precautions. If required,
the shaking head may be removed. First, disconnect the vortexer from the power source and then slowly
pull the shaking head straight up and off the shaking post. The head may then be cleaned with a mild
detergent. Be sure the head is thoroughly dry before replacing on the unit.
The motor and shaking mechanism in the Corning LSE Vortex Mixer requires no routine maintenance or
lubrication. To ensure a long life for the mixer, it should be used in a dry, non-corrosive environment.
Service and Repair Information
If you have a question about the Corning LSE
Vortex Mixer or have a service inquiry, contact Corning
Customer Service immediately at:
1.800.492.1110 (for toll-free calling within the U.S. and Canada)
1.978.442.2200 (outside the U.S.), or
Contact your local Corning support office listed in the back of this manual.
Before returning any unit for service, a Return Authorization (RA) number must first be obtained.
Equipment sent in without out prior authorization will be returned at the customer’s expense. When
returning a unit to Corning for service, it should be sent in the original packaging. If this is not possible,
be sure that the unit is sufficiently packed. Any damage resulting from improper packaging is the
responsibility of the customer. A written explanation should accompany the unit along with the RA
number.