WARNING:  Do not attempt to open the lid of any centrifuge until the rotor has 
come   to a complete stop. 

 

 

In the event of a power failure or malfunction, it may be necessary to open the 
lid manually.  

 

  1.   Disconnect the power cord from the wall socket.  

 

  2.   Remove the plastic plug, located on the left side of the unit.  

 

  3.   Pull the wire (attached to the plug) to open the lid lock manually.  

 

 

4.4  

Lid lock 

 

The centrifuge can be started only with the lid securely closed. When a run          
is started, the lid lock automatically activates. Do not attempt to open the lid 
during

 

a centrifuge run. At the end of the run, the lid will automatically unlock.  

Never attempt to override the lid lock mechanism. Doing so is dangerous 
and could damage the centrifuge.  

 

 

 

4.5  

Speed selection (see Figure 2.) 

 

The speed (rpm or g-force) can be selected from 500 to 13,300 rpm in 100 
rpm increments or from 100 to 16,300 x g with the knob on the right. “g-force” 
or   “rpm” display is set by pressing the knob. Turn the knob to increase or 
decrease the value.  

 

       

4.6  

Selection of operating time, momentary operation, Start/Stop 

 

Operating time can be selected from 0.5 minutes to 30 minutes by turning the 
timer knob on the left. The time can be set in 0.5 minute increments from 0 to 
10 minutes and in 1 minute increments from 10 to 30 minutes. After 30 
minutes, the display shows “--”  which indicates continuous run. In this mode, 
the centrifuge will run until manually stopped. 

To start a run, press the timer 

knob. 

 

When the preselected time expires, the centrifuge will stop automatically.          

To

 

stop

 

the centrifuge prior to the expiration of set time, press the timer 

knob. 

 

The centrifuge may be operated for a short run by pressing and holding the 
timer knob. The centrifuge will continue to run as long as the timer knob is 
depressed and the time, in seconds will count up on the time display.  

 

 

 
 

Summary of Contents for LSE 6765-HS

Page 1: ...INSTRUCTION MANUAL Models 6765 HS 120V US 6766 HS 230V EU 6767 HS 230V UK 6768 HS 100V JP CLS EQ 102 March 2010 Document 01 Revision 01...

Page 2: ...pletely stopped NEVER move the centrifuge while the rotor is spinning NEVER use solvents or flammables near this or other electrical equipment NEVER centrifuge flammable explosive or corrosive materia...

Page 3: ...nance Removing and installing the angle rotor Loading the rotor Overloading rotors Operation 5 Attaching rotor lid Closing the lid Lid release Lid lock Speed selection RPM g force Selection of operati...

Page 4: ...ng to operate the unit If this equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired DO NOT OPERATE the centrifuge IF any of the foll...

Page 5: ...d in the case of incorrect operation use of nonstandard spare parts or accessories and unauthorized modification of the rotor or centrifuge Corning reserves the right to make technical modifications P...

Page 6: ...d to the centrifuge and the power source 3 Installation of rotors and rotor maintenance 3 1 Rotors and accessories The following accessories are included or available for the Corning LSE High Speed Mi...

Page 7: ...screw counterclockwise using the rotor wrench Lift the rotor directly upward in a straight vertical motion To replace rotor first make sure the motor shaft and rotor mounting hole are clean Place the...

Page 8: ...lease contact the manufacturer for assistance 4 Operation ATTENTION Never attempt to operate the centrifuge with rotors or adapters that show signs of corrosion or mechanical damage Never centrifuge s...

Page 9: ...500 to 13 300 rpm in 100 rpm increments or from 100 to 16 300 x g with the knob on the right g force or rpm display is set by pressing the knob Turn the knob to increase or decrease the value 4 6 Sele...

Page 10: ...t maximum temperature 50 C 5 3 Cleaning the rotor The rotor should be cleaned after each use When spinning samples containing phenol or phenol chloroform the rotor should be cleaned immediately after...

Page 11: ...ossible reason Lid lock is jammed Solution Call for service Possible reason Centrifuge is not receiving power Solution See Centrifuge will not start Centrifuge cannot be started although power is on P...

Page 12: ...out out prior authorization will be returned at the customer s expense When returning a unit to Corning for service it should be sent in the original packaging If this is not possible be sure that the...

Page 13: ...is product fails within the specified period of time because of a defect in material or workmanship contact Corning s Customer Service at the following numbers USA 1 800 492 1110 Canada 1 978 442 2200...

Page 14: ...r a Class A digital device pursuant to both Part 15 of the FCC Rules and the radio interference regulation of the Canadian Department of Communications These limits are designed to provide reasonable...

Page 15: ...of with unsorted municipal waste It is the product user s responsibility to correctly dispose of waste equipment by handing it over to an authorized facility for separate collection and recycling It...

Page 16: ...4604000 t 31 20 655 79 28 f 91 124 4604099 f 31 20 659 76 73 Japan United Kingdom t 81 3 3586 1996 t 0800 376 8660 f 81 3 3586 1291 f 0800 279 1117 Korea t 82 2 796 9500 f 82 2 796 9300 All other Eur...

Reviews: