Solution: Call 

service 

 

Lid:  (Error message) 

Lid not closed 
 

Possible reason: 

Lid not closed completely 

Solution: Close 

lid 

 

Possible reason: 

Lid lock or sensor defective 

Solution: Call 

service 

 

Other Error Messages 

Er on Display 
 

Solution: Press 

time 

or 

speed knob to clear error 

 Call 

service 

  

7. Where to call 

 

If you have a question about the Corning

 LSE

 High Speed Microcentrifuge or have a 

service inquiry, contact Corning Customer Service immediately at: 

  1.800.492.1110 (for toll-free calling within the U.S. and Canada) 

 1.978.442.2200 

(outside 

the U.S.), or 

  Contact your local Corning support office listed in the back of this manual. 

 

Before returning any unit for service, a Return Authorization (RA) number must first be 
obtained.  Equipment sent in without out prior authorization will be returned at the 
customer’s expense.  When returning a unit to Corning for service, it should be sent in the 
original packaging.  If this is not possible, be sure that the unit is sufficiently packed.  Any 
damage resulting from improper packaging is the responsibility of the customer.  A written 
explanation should accompany the unit along with the RA number and a decontamination 
form. 

 

8. Determination of g-values 

 

The centrifuging radius of the 1.5/2.0 mL rotor is 8.23 cm. The Corning High Speed 
Microcentrifuge has an automatic g-force conversion program, so g-values are 
automatically calculated and can be displayed on the control panel, for this centrifuging 
radius. If adapters or smaller tubes are used, the centrifuging radius changes as does 
the g force.  

 

G- force can be calculated with a new measured radius and rpm using the following 
formula: 

 

 

g-force = 11.18 x r x 

(

 

)

2

 

 
 

 

 

r = radius in centimeters 

 

 

 

n = speed in rpm 

 

Summary of Contents for LSE 6765-HS

Page 1: ...INSTRUCTION MANUAL Models 6765 HS 120V US 6766 HS 230V EU 6767 HS 230V UK 6768 HS 100V JP CLS EQ 102 March 2010 Document 01 Revision 01...

Page 2: ...pletely stopped NEVER move the centrifuge while the rotor is spinning NEVER use solvents or flammables near this or other electrical equipment NEVER centrifuge flammable explosive or corrosive materia...

Page 3: ...nance Removing and installing the angle rotor Loading the rotor Overloading rotors Operation 5 Attaching rotor lid Closing the lid Lid release Lid lock Speed selection RPM g force Selection of operati...

Page 4: ...ng to operate the unit If this equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired DO NOT OPERATE the centrifuge IF any of the foll...

Page 5: ...d in the case of incorrect operation use of nonstandard spare parts or accessories and unauthorized modification of the rotor or centrifuge Corning reserves the right to make technical modifications P...

Page 6: ...d to the centrifuge and the power source 3 Installation of rotors and rotor maintenance 3 1 Rotors and accessories The following accessories are included or available for the Corning LSE High Speed Mi...

Page 7: ...screw counterclockwise using the rotor wrench Lift the rotor directly upward in a straight vertical motion To replace rotor first make sure the motor shaft and rotor mounting hole are clean Place the...

Page 8: ...lease contact the manufacturer for assistance 4 Operation ATTENTION Never attempt to operate the centrifuge with rotors or adapters that show signs of corrosion or mechanical damage Never centrifuge s...

Page 9: ...500 to 13 300 rpm in 100 rpm increments or from 100 to 16 300 x g with the knob on the right g force or rpm display is set by pressing the knob Turn the knob to increase or decrease the value 4 6 Sele...

Page 10: ...t maximum temperature 50 C 5 3 Cleaning the rotor The rotor should be cleaned after each use When spinning samples containing phenol or phenol chloroform the rotor should be cleaned immediately after...

Page 11: ...ossible reason Lid lock is jammed Solution Call for service Possible reason Centrifuge is not receiving power Solution See Centrifuge will not start Centrifuge cannot be started although power is on P...

Page 12: ...out out prior authorization will be returned at the customer s expense When returning a unit to Corning for service it should be sent in the original packaging If this is not possible be sure that the...

Page 13: ...is product fails within the specified period of time because of a defect in material or workmanship contact Corning s Customer Service at the following numbers USA 1 800 492 1110 Canada 1 978 442 2200...

Page 14: ...r a Class A digital device pursuant to both Part 15 of the FCC Rules and the radio interference regulation of the Canadian Department of Communications These limits are designed to provide reasonable...

Page 15: ...of with unsorted municipal waste It is the product user s responsibility to correctly dispose of waste equipment by handing it over to an authorized facility for separate collection and recycling It...

Page 16: ...4604000 t 31 20 655 79 28 f 91 124 4604099 f 31 20 659 76 73 Japan United Kingdom t 81 3 3586 1996 t 0800 376 8660 f 81 3 3586 1291 f 0800 279 1117 Korea t 82 2 796 9500 f 82 2 796 9300 All other Eur...

Reviews: