background image

10/16

ALLY II UPS™

12.5  REMOVING THE ALLY II UPS™

  1.  Detach the manipulator from the adapter drape by opening the latch per the Instructions for Use provided 

with the adapter. Detach the manipulator from the adapter drape prior to extracting the uterus during a  

hysterectomy procedure.

  2.  If used, undock the da Vinci

®

 Surgical System and move patient side cart away from the operating room table.

  3.  Detach the adapter by squeezing the pinch clips together.

  4.  Remove the used sterile adapter drape and dispose of properly.

NOTE: 

Refer to the adapter drape’s Instructions for Use for complete instructions.

CAUTION:

  

The flexible arm must be rotated out of the way or the ALLY II UPS™ MUST be removed from the 

table PRIOR to the foot end/pins being returned to the horizontal position.

12.6  POWERING DOWN THE ALLY II UPS™

  1.  Press the foot pedal while powering down so that the arm remains in its flexible state when dismounting and storing. 

Operating room staff may choose to store the flexible arm in a rigid position as appropriate.

  2.  Turn off the power switch and unplug from the wall outlet.

  3.  Fully loosen the knob on the operating room table rail camp until the ALLY II UPS™ is no longer firmly secured 

to  the operating room table. The ALLY II UPS™ can be lifted off the rail clamp. Figure 4 shows how to remove 

the ALLY II UPS™ main unit from the rail.

  4.  Remove the rail clamp and store it with the ALLY II UPS™.

12.7  CLEANING AND STORAGE

  1.  The ALLY II UPS™ system including main unit, fixed arm, handle and foot pedal should be cleaned after each 

use. Only use Isopropyl Alcohol 70 percent (70% or higher) or CaviWipes™ (or equivalent) not to exceed .28% 

ammonium chloride. Wipe all surfaces free of debris. Do not spray flexible arm. Ensure flexible arm is locked in 

upright orientation during cleaning application. Ensure flexible arm is thoroughly dry before powering on.

  2.  After each use clean the rail clamp; clean and disinfect using a quaternary ammonium disinfecting/cleaning 

solution, read the cleaning product’s directions and follow the instructions on the label for recommendation for 

achieving low-level disinfection; use caution around the knob where fluid migration may occur. Wipe device with 

a clean, dry cloth; make certain the product is dry prior to reinstalling to avoid damage.

CAUTION:

  

Damage will result if the flexible arm is cleaned with bleach products.

CAUTION:

  

Damage may result if the rail clamp is cleaned with bleach.

CAUTION:

  

Care should be taken during storage and transportation to avoid drops, falls, mechanical stress, 

and mechanical shock to the device.

  3.  The foot pedal cover should be used to lift, carry or reposition the foot pedal.

  4.  Store this device in a clean, dry and well-ventilated environment.

WARNING:

  

DO NOT

 store the device in direct sunlight, at high temperatures or high humidity.

WARNING:

  

DO NOT

 store this device in the shipping box.

Figure 4:

 Removing the ALLY II UPS™ main unit from the rail

Summary of Contents for ALLY II UPS

Page 1: ...ning System Instructions for Use AU UPS2 CLP4274203 Rev A 11 22 English en 95 Corporate Drive Trumbull CT 06611 USA Phone 800 243 2974 Fax 800 262 0105 International Phone 1 203 601 9818 Fax 1 203 601 4747 www coopersurgical com ...

Page 2: ...3 ATTACHING THE ADAPTER DRAPE 9 12 4 INTRA OPERATIVE USE 9 12 5 REMOVING THE ALLY II UPS 10 12 6 POWERING DOWN THE ALLY II UPS 10 12 7 CLEANING AND STORAGE 10 SECTION 13 GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS for all Medical Equipment and Medical Equipment Systems 11 SECTION 14 GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC IMMUNITY for all Medical Equipment and...

Page 3: ...al movement from a single point allowing the system to be attached to one side of the operating room table Foot pedal release facilitates positioning Lock arm in the desired position by releasing the foot pedal Connects to any operating table Provides static control consistent cephalad pressure and stability SECTION 7 CLINICAL BENEFIT Holds the uterus in the desired position and allows for visuali...

Page 4: ...e separation between the interfering equipment and the ALLY II UPS Incrementally turn off equipment in the vicinity to identify the interfering device PRECAUTIONS Always ensure that the ALLY II UPS is tightly and securely attached to the table prior to the start of the surgical procedure Improper or loose mounting of the system can lead to unintended movement and may lead to injury Inspect the ada...

Page 5: ...ipulator in a desired position The ALLY II UPS consists of the ALLY II UPS and the manipulator adapter with built in sterile drape sold separately known as the adapter drape Figure 1 shows the system followed by descriptions of its various parts Figure 1 ALLY II Uterine Positioning System ALLY II UPS NON STERILE Must drape prior to use as shown Manipulator Adapter Sterile Drape Flexible Arm ALLY I...

Page 6: ...lling or removing the ALLY II UPS from the rail Refer to Section 12 2 for more information Power Cord A detachable power cord is included which supplies power to both the main unit and the foot pedal Grounding Post To access ground connection 2 ALLY II UPS RAIL CLAMP Used to mount the ALLY II UPS to the patient s right side rail of standard operating room tables Secure by tightening the knob 3 ADA...

Page 7: ...urer s instructions for proper use and positioning Ensure patient is placed such that the buttocks are as close to edge as possible 2 Lower or remove foot end of the operating room table 3 Attach the ALLY II UPS to the operating room table by placing the rail clamp on the patient s right side near the stirrup Lower the mounting tab of the ALLY II UPS into the rail clamp 4 Tighten the knob by hand ...

Page 8: ...sonnel actuating theALLY II UPS It is recommended to position the ALLY II UPS foot pedal separately from other foot pedals CAUTION Securely hold arm or manipulator before depressing foot pedal 7 Turn the power on and verify that the green light indicator is lit Press and hold the foot pedal once Release to initialize the ALLY II UPS The ALLY II UPS arm can be moved by grasping the flexible arm and...

Page 9: ...flexible arm during the remaining patient preparation steps 3 It is recommended to drape the patient s legs at this step in the process NOTE Continue with necessary patient preparation steps 12 4 INTRA OPERATIVE USE 1 Grasp the handle of the manipulator or the ALLY II UPS flexible arm 2 Depressing the foot pedal allows for dynamic manipulation of the flexible arm 3 Maneuver to the desired position...

Page 10: ...the rail clamp and store it with the ALLY II UPS 12 7 CLEANING AND STORAGE 1 The ALLY II UPS system including main unit fixed arm handle and foot pedal should be cleaned after each use Only use Isopropyl Alcohol 70 percent 70 or higher or CaviWipes or equivalent not to exceed 28 ammonium chloride Wipe all surfaces free of debris Do not spray flexible arm Ensure flexible arm is locked in upright or...

Page 11: ... GHz b 80 AM at 1 kHz c 10 V m f 80 MHz 2 7 GHz b 80 AM at 1 kHz c Proximity fields from RF wireless communications equipment IEC 61000 4 3 See 8 10 IEC 60601 1 2 Ratedpowerfrequencymagnetic fields d e IEC 61000 4 8 30 A m g 50 Hz or 60 Hz a The interface between the patient physiological signal simulation if used and the ME equipment or ME system shall be located within 0 1 m of the vertical plan...

Page 12: ...r input of the converter g Applicable only to ME equipment and ME systems connected to single phase a c mains h E g 10 12 means 10 periods at 50 Hz or 12 periods at 60 Hz i ME equipment and ME systems with rated input current greater than 16 A phase shall be interrupted once for 250 300 cycles at any angle and at all phases at the same time if applicable ME equipment and ME systems with battery ba...

Page 13: ...lation b 217 Hz 2 0 3 28 5 240 5 100 5 800 WLAN 802 11 a n Pulse Modulation b 217 Hz 0 2 0 3 9 5 500 5 785 NOTE If necessary to achieve the immunity test level the distance between the transmitting antenna and the ME equipment or ME system may be reduced to 1 m The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be mo...

Page 14: ... the date of purchase If CooperSurgical Inc determines that the ALLY II UPS fails to perform within that one year as the sole remedy for that failure to perform at CooperSurgical s discretion we will repair or replace the ALLY II UPS free of charge SECTION 19 SERVICE Annual service by authorized CooperSurgical personnel is required to ensure proper device operation Please contact CooperSurgical Re...

Page 15: ... Slow Blow Classification I IP Rating Main Unit IP30 Foot Pedal IPX6 Average Duty Cycle Rating 5 seconds ON 300 seconds OFF No customer replaceable components inside the main unit assembly Environmental Conditions Operational Temperature 68 F to 75 F 20 C to 24 C Humidity 20 RH to 60 RH Shipping and Storage Temperature 14 F to 122 F 10 C to 50 C Humidity 10 RH to 85 RH 7 5 19 cm 15 4 39 0 cm 22 3 ...

Page 16: ...erSurgical is a registered trademark of CooperSurgical Inc 2022 CooperSurgical Inc Symbol Title Manufacturer Caution Catalogue number Serial number Country of manufacture Date of manufacture Consult instructions for use 6D33 UL Rating Approved MEDICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK FIRE AND MECHANICAL HAZARD ONLY IN ACCORDANCE WITH ANSI AAMI ES60601 1 2005 C1 2009 A2 2010 and CAN CSA C...

Reviews: