SECTION 8 - CARE & MAINTENANCE

Immersion in media solutions will accelerate corrosion of the probe 

tip due to the salt content of the media. Similarly, long-term exposure 

tohumidifiedairinsideanincubatormayacceleratecorrosion.To

prolong the life of the probe, we suggest using plain water drops 

instead of culture media when checking the temperature in a Petri 

dish (see Media Management on page 9)
After each use, rinse the probe tip in plain water, blot dry gently and 

allow to air dry. Then replace the plastic cap before storing.

Calibration

We recommend calibrating the IVF Thermometer annually. This can 

be done by RI, but it may be more convenient for most users to use 

a local test company. We recommend calibration be conducted by 

a test house accredited to national standards. The meter should be 

calibratedtothespecificationsshownon

page 13

Calibrationdoesnotinvolveadjustingthemeter,rathercomparison

to a traceable standard offset. For example, if the calibration

certificateshowsareadingof37.3°Catanactual37.0°Cthenthe

meterwilldothisconsistently.Thisoffsetcanbeallowedforwhen

making measurements.
Replacing a probe always requires a new calibration. Similarly, 

the meter can be used with more than one probe, but each probe 

should be calibrated separately with the meter, and care should be 

takentoapplythecorrectoffsetforeachprobe.

The cryo-probe does not require calibration as the inherent 

accuracyissufficientforthepurpose.Seethespecificationtableon

the following page.

Summary of Contents for 2110-T

Page 1: ...2 02 customerservice origio com fertility coopersurgical com Research Instruments Ltd Bickland Industrial Park Falmouth Cornwall TR11 4TA UK Document 7 90 000UM 4 DRF 5311 6 February 2020 RI Thermomet...

Page 2: ......

Page 3: ...ction 2 Applicable Part Numbers 2 SECTION 3 SAFETY WARNINGS 3 Glossary of Safety and Information Symbols 3 SECTION 4 PRODUCT OVERVIEW 4 Inventory 4 SECTION 5 SET UP 5 Batteries 5 Probe 7 SECTION 6 BAS...

Page 4: ...NTS SECTION 7 TROUBLESHOOTING 1 1 SECTION 8 CARE MAINTENANCE 12 Calibration 12 Specification 13 SECTION 9 WARRANTY INFORMATION AND LIMITS ON LIABILITY 14 SECTION 10 RETURNING PRODUCT TO RI FOR REPAIR...

Page 5: ...used for Assisted Reproductive Technology ART procedures Contraindication It is not intended that this device come into direct contact with any tissue samples being treated in a clinical setting or w...

Page 6: ...aboratories use the IVF Thermometer to verify that any temperature controlled equipment is performing as desired The unit is waterproof and is calibrated for use with refrigerators incubators and heat...

Page 7: ...onsult instructions for use coopersurgical com Hard copies available on request Indicates the need for the user to consult the instructions for use for important cautionary information such as warning...

Page 8: ...be Temperature Probe stand Thermometer and 2 Probes IVF Thermometer 2 Probes Temperature Probe stand ON and OFF buttons HOLD button Display screen Repositionable tape Thermal transfer compound 2ml syr...

Page 9: ...Made in England AA Read Instructions Before Use Type T Thermocouple Thermometer Range 250 C to 400 C 418 F to 752 F Screws SECTION 5 SET UP Batteries The IVF Thermometer is powered by two AA batteries...

Page 10: ...e 250 C to 400 C 418 F to 752 F Research Instruments Ltd Tel 44 0 1326 372 753 www research instruments com Made in England AA 2 Fit the batteries as shown The orientation of the batteries must be as...

Page 11: ...opreservation temperatures Connect the probe to the socket at the top of the meter and tighten the screw collar Remove the white plastic sleeve from the probe tip before use but keep the sleeve to pro...

Page 12: ...on the display Press the HOLD button again to return to the live display To switch the meter off press and hold the OFF button The meter will also automatically switch off after about five minutes if...

Page 13: ...tip can be immersed directly in the water drop Allow the reading to settle before recording or adjusting the temperature The probe can be held by hand or placed in the holder supplied Incubator Measu...

Page 14: ...red 2 Insert the tip of the probe into the paste 3 Cut a short piece of the adhesive tape provided and secure the probe to the surface by placing the tape lengthways as shown If the tape is placed acr...

Page 15: ...Research Instruments Ltd 7 SECTION 7 TROUBLESHOOTING Display shows O C open circuit this means the probe has failed The most common cause of failure is a broken wire at the tip Failed probes cannot be...

Page 16: ...be more convenient for most users to use a local test company We recommend calibration be conducted by a test house accredited to national standards The meter should be calibrated to the specification...

Page 17: ...m Weight of units 250g Probe Type Optional robust probe Temperature range 0 to 100 C Accuracy calibrated 0 2 C 0 4 F over the range 20 C to 70 C 4 F to 158 F 0 1 of reading 0 2 C 0 4 F elsewhere which...

Page 18: ...nties do not cover damage caused by misuse improper care improper use of chemicals or cleaning methods loss theft use of non authorized parts servicing by non authorized personnel or negligent or inte...

Page 19: ...r RI shall not under any circumstances be liable for lost profits damages from loss of use or lost data or indirect special incidental or consequential damages under its warranties or otherwise for an...

Page 20: ...ccepted without an RMA number Please have the following information Reason for returning the goods Quantity description part number serial number of the goods Date of receipt of order Customer s purch...

Page 21: ...ter 60 days No refund on sterile single use disposable products Customer Service Contact details Tel 45 46 79 02 02 Fax 45 46 79 03 02 E mail customerservice origio com fertility coopersurgical com Co...

Page 22: ......

Page 23: ......

Page 24: ......

Reviews: