COOK Medical RETRACTA Instructions For Use Manual Download Page 10

10

spin echo and gradient echo pulse sequence in a MR 3.0 Tesla system (General 
Electric Healthcare Excite, HDx).  Therefore, it may be necessary to optimize 
MR imaging parameters for the presence of this metallic embolization coil.

For US Patients Only

Cook recommends that the patient register the MR conditions disclosed in 
this IFU with the MedicAlert Foundation. The MedicAlert Foundation can be 
contacted in the following manners:

Mail: 

MedicAlert Foundation International 

2323 Colorado Avenue 

Turlock, CA 95382

Phone:  888-633-4298 (toll free) 

209-668-3333 from outside the US

Fax: 

209-669-2450

Web: www.medicalert.org

INSTRUCTIONS FOR USE

1.  Perform an angiogram and measure the diameter of the vessel to be 

occluded.

2.  Firmly grasping the loading cartridge between thumb and forefinger, 

introduce the metal end of the loading cartridge into the base of the 
catheter hub. Lock the loading cartridge onto the catheter hub by turning 
the Luer lock adapter clockwise. (Fig. 3)

3.  Unlock the white safety clip and advance at least 30 cm of the delivery 

wire into the catheter so that the coil is fully loaded in the catheter.  
(Fig. 4)

4.  Unlock the loading cartridge from the catheter hub by turning the Luer 

lock adapter counter-clockwise.

5.  Remove the delivery wire holder and the metal loading cartridge from 

the catheter hub while holding the delivery wire stationary. Remove the 
torque device and reserve it for use later in the procedure.

6.  Under fluoroscopic visualization, slowly advance the delivery wire until 

the entire length of the coil exits the distal end of the catheter. Ensure 
that the junction remains positioned just inside the catheter tip. (Fig. 5)

  NOTE: Advancing the delivery wire slowly allows the junction to be seen 

more easily and reduces the risk of damaging it.

  NOTE: If significant resistance is encountered during coil advancement, 

do not continue advancing. Retract the delivery wire slightly, then gently 
re-advance it. If there is still significant resistance, withdraw the delivery 
wire from the catheter and try using a new coil with a shorter length.

  NOTE: Do not turn the delivery wire counterclockwise during 

advancement; the coil may be unintentionally detached.

7.  Verify correct position of the coil fluoroscopically. If coil position or 

placement is not satisfactory, the coil may be retracted into the catheter 
and re-deployed so long as there is no significant resistance.

  NOTE: It may be possible to perform a test injection of contrast media 

using a Tuohy-Borst Sidearm Adaptor while the delivery wire is in the 
catheter.

  NOTE: If the size of the coil is not correct, gently remove the entire 

delivery wire and coil. Do not use the coil again.

8.  If the coil position is correct, use the torque device to turn the delivery 

wire counterclockwise 8-10 times, until coil detachment can be either felt 
or visualized under fluoroscopy. (Fig 6)

  NOTE: It is recommended that the junction remain just inside the tip of 

the catheter.

  NOTE: Do not advance the delivery wire after the coil is detached.
9.  Gently remove the delivery wire after coil detachment.

10.  Additional Retracta or pushable coils may be required to achieve 

permanent occlusion of the vessel.

Coil Delivery Technique and Coil Size Selection

Long-term occlusion depends on achieving cross sectional occlusion of the 
blood vessel, and coaxial catheters provide the ability to control placement 
of coils and permanent occlusion. The combination of the coaxial technique 
and either the anchor or scaffold technique significantly enhances stability of 
coil deployment.

•  Coaxial technique: The use of outer guiding sheath/catheter is the most 

important step in preventing coil elongation and uncertain long-term 
occlusion. The outer guiding sheath/catheter provides support, and the 
inner catheter provides finer selective maneuvers. (Fig. 7)

•  Anchor technique: The anchor technique provides safe and distal 

occlusion when there is a question about instability of coils. At least 2 cm 
of a coil is advanced into the side branch, which is normally sacrificed. 
The rest of the coil is then deployed just proximal to that side branch, and 
additional coils are packed. (Fig. 8)

•  Scaffold technique: The scaffold technique is used for high-flow vessels 

when there is concern about migration of a softer coil. A high radial force 
coil is placed initially. Then, several high radial force coils or soft coils may 
be packed within the scaffold. (Fig. 9)

Summary of Contents for RETRACTA

Page 1: ...Retracta Instrucciones de uso Spirales d embolisation d tachables Retracta Mode d emploi Spirali per embolizzazione a rilascio controllato Retracta Istruzioni per l uso Retracta loskoppelbare embolisa...

Page 2: ...Luer 1 Clip de seguridad 2 Cartucho de carga 3 Dispositivo de torque 4 Adaptador Luer Lock 1 Clip de s curit 2 Cartouche de chargement 3 Dispositif de contr le de torque 4 Adaptateur Luer Lock 1 Clip...

Page 3: ...Coil 1 Fremf ringstr d 2 Forbindelsespunkt 3 Coil 1 Platzierungsdraht 2 bergangsbereich 3 Spirale 1 2 3 1 Alambre de implantaci n 2 Uni n 3 Espiral 1 Guide de largage 2 Jonction 3 Spirale 1 Spingitore...

Page 4: ...Illustrations Illustrationer Abbildungen Ilustraciones Illustrations Illustrazioni Afbeeldingen Ilustra es Illustrationer 3 4 5 6...

Page 5: ...ath 3 Catheter 1 Snoning 2 Sheath 3 Kateter 1 Hin und Herbewegung 2 Schleuse 3 Katheter 1 2 3 1 Zigzagueo 2 Vaina 3 Cat ter 1 Mouvement de va et vient circulaire 2 Gaine 3 Cath ter 1 Movimento alterna...

Page 6: ...ath 3 Catheter 1 Snoning 2 Sheath 3 Kateter 1 Hin und Herbewegung 2 Schleuse 3 Katheter 1 2 3 1 Zigzagueo 2 Vaina 3 Cat ter 1 Mouvement de va et vient circulaire 2 Gaine 3 Cath ter 1 Movimento alterna...

Page 7: ...coil med stor radial kraft stillads 6 Platin coil bl d coil 7 Anlagt platin coil tv rsnitsokklusion gennemf rt 1 Hin und Herbewegung 2 Schleuse 3 Katheter 4 Spirale mit hoher Radialkraft 5 Platzierte...

Page 8: ...a 7 Spirale in platino rilasciata occlusione della sezione trasversale comple tata 1 Wevende beweging 2 Sheath 3 Katheter 4 Coil met grote radiale kracht 5 Coil met grote radiale kracht scaffold gepla...

Page 9: ...ization to determine correct catheter position Prior to introduction of the embolization coil flush the angiographic catheter with saline This product is intended for use by physicians trained and exp...

Page 10: ...wire counterclockwise during advancement the coil may be unintentionally detached 7 Verify correct position of the coil fluoroscopically If coil position or placement is not satisfactory the coil may...

Page 11: ...re anbragte coils Der skal opretholdes en minimal men tilstr kkelig arteriel blodgennemstr mning for at fastholde coils ene mod de tidligere anbragte coils indtil en fast koagulation sikrer permanent...

Page 12: ...Fjern holderen til fremf ringstr den og ladehylsteret af metal fra katetermuffen samtidig med at fremf ringstr den holdes ubev gelig Fjern momenth ndtaget og l g det til side til brug senere hen i in...

Page 13: ...tracta abl sbare Embolisationsspirale ist f r die Einbringung in das Zielgef unter Durchleuchtung konzipiert Mit dem System der Retracta abl sbaren Embolisationsspirale kann diese vor der endg ltigen...

Page 14: ...xcite HDx Es kann daher erforderlich sein die Parameter der MRT Bildgebung aufgrund des Vorliegens dieser metallischen Embolisationsspirale zu optimieren Nur f r Patienten in den USA Cook empfiehlt da...

Page 15: ...ieten die F higkeit die Platzierung der Spiralen und die permanente Okklusion zu steuern Mit einer Kombination aus der koaxialen Technik und entweder der Anker oder der Ger sttechnik l sst sich die St...

Page 16: ...NB R 4 0 35 HNB R 5 0 35 HNB R 5 0 38 SCBR 5 0 38 SCBR 4 0 38 PTBYC RA Retracta Retracta ASTM F2503 3 0 Tesla 1 600 Gauss cm SAR 2 0 W kg 15 Retracta 1 1 C SAR 2 0 W kg 15 3 0 Tesla General Electric E...

Page 17: ...ook MedicAlert Foundation MedicAlert Foundation MedicAlert Foundation International 2323 Colorado Avenue Turlock CA 95382 1 888 633 4298 1 209 668 3333 1 209 669 2450 www medicalert org 1 2 Luer 3 3 3...

Page 18: ...ito ADVERTENCIAS La colocaci n de espirales de embolizaci n debe llevarse a cabo con especial cuidado Las espirales no deben dejarse demasiado cerca de las entradas de arterias y si es posible deben e...

Page 19: ...agen La calidad de la imagen de la MRI puede resultar afectada si la zona de inter s est a menos de unos 9 mm de la posici n de la espiral de embolizaci n largable Retracta como se observ durante prue...

Page 20: ...vaso sangu neo y los cat teres coaxiales permiten controlar la colocaci n de las espirales y la oclusi n permanente La combinaci n de la t cnica coaxial y la t cnica bien de anclaje o bien de andamio...

Page 21: ...utilis par des m decins d ment form s et exp riment s dans les techniques d embolisation art rielle et veineuse Le m decin proc dera selon les m thodes classiques de pose de gaines d acc s vasculaire...

Page 22: ...l avancer nouveau avec pr caution S il existe toujours une r sistance significative retirer le guide de largage du cath ter et essayer d utiliser une nouvelle spirale avec une longueur plus courte RE...

Page 23: ...sotto fluoroscopia Il sistema della spirale per embolizzazione a rilascio controllato Retracta ne permette l avanzamento e il completo riposizionamento prima del rilascio finale della spirale USO PREV...

Page 24: ...ametri di imaging MRI considerando la presenza di questa spirale metallica per embolizzazione Solo per pazienti negli USA Cook consiglia al paziente di comunicare alla MedicAlert Foundation le condizi...

Page 25: ...avanzare almeno 2 cm di una spirale nella diramazione laterale che normalmente viene sacrificata Il resto della spirale viene quindi rilasciato in posizione appena prossimale rispetto a tale diramazi...

Page 26: ...coil bij het toedienen van een testinjectie met contrastmiddel MRI INFORMATIE Niet klinisch onderzoek heeft aangetoond dat de Retracta loskoppelbare embolisatiecoil onder bepaalde voorwaarden MRI veil...

Page 27: ...Borst zijarmadapter terwijl de plaatsingsdraad zich in de katheter bevindt NB Als de afmetingen van de coil niet juist zijn moeten de hele plaatsingsdraad en de coil voorzichtig worden verwijderd Gebr...

Page 28: ...deixar as espirais demasiado perto da entrada das art rias e se poss vel as espirais devem ser entrecruzadas com espirais previamente colocadas Com o objectivo de manter as espirais contra as espirais...

Page 29: ...roduzir a extremidade de metal do cartucho de carregamento na base do conector do cateter Fixar o cartucho de carregamento ao conector do cateter rodando o adaptador Luer Lock no sentido dos ponteiros...

Page 30: ...liza o Est ril desde que a embalagem n o esteja aberta nem danificada Se tiver alguma d vida quanto esterilidade do produto n o o utilize Guarde num local protegido da luz seco e fresco Evite a exposi...

Page 31: ...f rdes i ett 3 0 teslas MR system General Electric Excite HDx programversion 14X M5 Bildartefakt MR bildkvaliteten kan vara nedsatt om omr det av intresse befinner sig inom cirka 9 mm fr n positionen...

Page 32: ...val av spiralstorlek L ngsiktig ocklusion r beroende av att man uppn r genomg ende ocklusion av blodk rlet och koaxialkatetrar ger m jlighet att kontrollera placeringen av spiraler och g ra ocklusione...

Page 33: ......

Page 34: ......

Page 35: ......

Page 36: ...MR Conditional MR conditional Bedingt MRT kompatibel MR conditional MR conditional compatible avec l irm sous certaines conditions Pu essere sottoposto a MRI MRI veilig onder bepaalde voorwaarden MR c...

Reviews: