1-7
K-MINC-1000 Instructions for Use
English
3.4 Rear of the device
Description
1.
Mains power inlet
Connect the appropriate power cord to this point.
2.
Gas inlet
Connect gas supply to this point.
3.
Gas outlet
Blanking plug shown.
4.
External alarm contacts
Connect to external alarm if required.
5.
USB connector
Type B socket provided.
External alarm connection:
Use only with suitable alarm transmitters activated by means of contact closure. Battery or Safety Extra Low Voltage
powered alarm equipment, not exceeding stated contact ratings may be connected to the external alarm terminals.
Refer to Technical Specifications (§8) for alarm contact rating.
USB connection:
Use only with Laptop or Desktop PC standard USB connections (USB 1.1 and USB 2.0 compliant).
Gas connections:
Refer to Gas Supply (§ 3.7) for details.
3.5 Electromagnetic compatibility
The MINC Mini Incubator has been tested and found to comply with the electromagnetic compatibility (EMC) limits
for medical devices as specified by IEC 60601-1-2:2014. These limits are designed to provide reasonable protection
against harmful interference in a typical medical installation.
Medical electrical equipment requires special precautions regarding EMC and must be installed and
operated according to these instructions. It is possible that high levels of radiated or conducted radio-frequency
electromagnetic interference (EMI) from portable and mobile RF communications equipment or other strong or
nearby radio-frequency sources could result in performance disruption of the MINC. Evidence of disruption may
include erratic readings, equipment ceasing to operate, or other incorrect functioning. If this occurs cease using the
MINC and contact your Cook Medical authorised service agent.
For guidance and manufacturer’s declaration on electromagnetic emissions and immunity of the MINC, refer to § 8.
WARNING: ELECTRIC SHOCK
HAZARD.
Determine if the available
voltage corresponds to the device.
Connecting to the wrong voltage will
cause the MINC to malfunction or may
permanently damage the device!
The power cord must be equipped with
a safety plug. Use the enclosed power
cord for the connection between the
power plug and the device socket!
WITHIN CANADA & U.S.A – Use only a
listed detachable power supply cable,
type SJT, minimum 18AWGx30,
3 conductors, one end configured
for NEMA 5-15, other end for
IEC 320/CEE22!
To avoid the risk of electric shock, this
equipment must only be connected to
a supply mains with protective earth.
1
3
4
2
5
1