Installation
November 8 2019 | 291180-ENG R00 13
3.5 Airflow
Continuous airflow is critical to the efficient operation of the chamber. Fresh air may be
introduced using the fresh air port located on the side of the chamber.
Conditioned air is drawn down from the refrigeration coil and enters the interior of the chamber
thought a rigid metal plenum in the rear wall of the chamber. Air is then directed horizontally
from the rear plenum, providing uniform environmental conditions.
3.6 Adjustable Wire Shelves
Wire shelves are supported by two metal support brackets attached to the four pilaster strips
inside the chamber (Figure 3-5). The chamber comes with five wire shelves to be spaced
evenly. An additional 10 shelves may be ordered to expand the shelving area to 15 shelves with
a minimum gap between shelves of 3
″ (7.62cm). Optional seedling trays (Figure 3-5) may also
be ordered.
Figure 3-5
Wire Shelves, Support Brackets, and Seedling Tray
3.6.1 Adjustable Shelf Spacing
Shelf configurations with up to 10 shelves:
•
Install the first shelf at 10” (25.4cm) below the chamber ceiling.
•
Install all remaining shelves
with a 4” (10.16cm) space between shelves.
Shelf configurations with between 10 and 14 shelves:
•
Install the first shelf at 10” (25.4cm) below the chamber ceiling.
•
Install all remaining shelves with a
3” (7.62cm) space between shelves.
Shelf configurations with 15 shelves:
•
Install the first shelf at 7” (17.78cm) below the chamber ceiling.
•
Install a
ll remaining shelves with a 3” (7.62cm) space between shelves.
Summary of Contents for GEN1000-GE
Page 1: ...GEN1000 GE OPERATOR MANUAL 291180 ENG R00...
Page 12: ...GEN1000 GE Operator Manual 4 November 8 2019 291180 ENG R00...
Page 16: ...GEN1000 GE Operator Manual 8 November 8 2019 291180 ENG R00...
Page 26: ...GEN1000 GE Operator Manual 18 November 8 2019 291180 ENG R00...
Page 34: ...GEN1000 GE Operator Manual 26 November 8 2019 291180 ENG R00...
Page 46: ...GEN1000 GE Operator Manual 38 November 8 2019 291180 ENG R00...
Page 53: ......