EMA-D9 - USER MANUAL
-52-
11) PROBLEMS
AND
SOLUTIONS
If you have a problem setting up or using your instrument, you may be able to solve it yourself.
Before calling your retailer or nearest distributor you should try the suggested actions that are
appropriate to your problem.
Problem Possible
cause
Suggested
The instrument doesn’t turn on.
- The power supply is
disconnected or wrong.
- The internal fuse is interrupted.
- Verify the connection and the
presence of power supply.
- See the chap. cap. 6.1) to
verify and/or to change the
internal fuse.
The display is completely dark
or clear.
- Contrast badly adjusted
- Adjust contrast as described
on 8.1)
The instrument doesn’t
communicate with the NRG
software (or other
communication software).
- Communication wires.
- Communication protocol.
- Wiring system and
communication parameters.
- Verify the correct wiring.
- Verify that the communication
protocol of the instrument
coincides with the one used in
the sw.
- Verify the wiring type (RS232
or RS485) and the settings of
the serial port of the instrument.
The instrument communicates
with the PC but the
communication is interrupted.
- Not shielded wires.
- Lack of terminations.
- Use shielded wires.
- Connect terminations as par.
7.4.1) and 7.4.2)
The instrument loses the CODE
or the PASSWORD.
- E
2
prom lost data.
- to try again to input another
time the data lost.
If the problem have not been solved, or for other information not covered in the present manual,
please contact with our Technical Assistance Department.
Before contacting, it is suggested to collect the maximum information regarding the installation, and
mainly the following data:
1. Model and serial number from the label on the top of the instrument housing.
2. Purchase
receipt.
3. Description
of
problem.
4. System configuration (hardware fitted, firmware release etc.).