414L-M-V3
8.2 WIRING DESIGNATIONS FOR THE MODEL 414L
Terminal
Model 414LR
Model 414LA
1
Calibration Link
Calibration Link
2
Signal Ground
Signal Ground
3
Flow Pulse Input (Channel 2)
Flow Pulse Input (Channel 2)
4
PT100 I (+)
Not To Be Used
5
PT100 Signal (+)
Temperature Input (4-
20mA)
6
PT100 Signal (-)
Not To Be Used
7
Flow Alarm
Flow Alarm
8
Flow Common (-)
Flow Common (-)
9
Flow Pulse Input (Channel 1)
Flow Pulse Input (Channel 1)
10
Pulse Out
Pulse Out
11
DC Power Out (8-24 VDC)
DC Power Out (8-24 VDC)
12
DC Ground
DC Ground
13
DC Power Input
DC Power Input
14
PT100 I (-)
Flow Input (4-20mA)
Terminal
RS232/422 /485 Option
20
RS232 Signal Ground
21
RS232 Data In
22
RS232 Data Out
23
RS422/485 (-) Data Out
24
RS422/485 (+) Data Out
25
RS422/485 (-) Data In
26
RS422/485 (+) Data In
27
RS232 CTS
Terminal
Relay Output Switches
2B
Signal Ground
28
Remote RUN Switch
29
Remote STOP Switch
30
End of Batch
31
Relay 2 - Normally Open
32
Relay 2 - Normally Closed
33
Relay 2 - Common
34
Relay 1 - Normally Open
35
Relay 1 - Normally Closed
36
Relay 1 - Common
Installation
73