Contec CMS800G User Manual Download Page 26

The   sterilization   the   manufacturer   recommended   to   cleaning   the   monitor   and   accessories   are 
ethanol and Acetaldehyde.

CAUTION: To avoid damaging the monitor:

1

Follow the  manufacturer’s instruction to  dilute the  solution, or adopt  the lowest possible 
density.

2

Do not let liquid enter the monitor.

3

No part of this monitor can be subjected to immersion in liquid.

4

Do not pour liquid onto the monitor during sterilization.

5

Wipe the device with a clean moistened cloth to remove any remaining sterilant.

6.7 Disinfection

To avoid extended damage to the equipment, disinfect ion is only recommended when stipulated 
as necessary in the hospital maintenance schedule. Disinfections facilities should be cleaned first.

CAUTION: 

 Follow the manufacturer’s instruction to dilute the solution.

 Do not use bleaching powder containing chloros on any parts of the monitor.

 Do not disinfect the monitor and it’s accessories with autoclave④gassing④formaldehyde process 

or radiation.

  Check carefully after cleaning, sterilization, or disinfect ion of monitor and accessories. If 

aging and damage are found, please do not use them to monitor.

Note

:   The   manufacturer   has   no   responsibility   for   the   effectiveness   of   controlling   infectious 

disease   using   these   chemical   agents.   Please   contact   infectious   disease   experts   in   your 
hospital for details.

7 Warranties

The manufacturer warrants that the Fetal Monitor we sell is free from defects in material and 
workmanship. In the status of normal operation and maintenance, if the manufacturer receives 

notice   of   such   defects   during   the   warranty   period   that   begins   on   the   date   of   shipment,   the 
manufacturer shall, at its options, either repair or replace hardware products that prove to be 

defective. The unit is guaranteed for periods of 12 months, valid from the date of purchase. The  
manufacturer also provides long-term repair service for our clients.

The manufacturer’s obligations or liability under this warranty does not include any transportation 
or other charges or liability for direct, indirect or consequential damages or delay resulting from 

the following conditions.

The following conditions are not included in the warranty:

1

Assembly operation, extensions, re-adjustments are carried out by the importer.

2

Application of the products or repaired by anyone other than the manufacturer authorized 
representative.

3

This warranty shall not extend to any instrument that has been damaged subjected to misuse, 

Summary of Contents for CMS800G

Page 1: ...CONTEC MEDICAL SYSTEMS CO LTD CMS800G FETAL MONITOR...

Page 2: ...er and Removing Paper Jam 17 4 5 Power On the Monitor 17 4 6 Connect Transducers 17 5 Monitoring 17 5 1 Operation Procedure 17 5 2 Print Operation 20 5 3 Operation After Monitoring 21 6 Maintenance Ca...

Page 3: ...nstallation of the relevant room complies with national standards and the instrument is used in accordance with the instructions for use Note This device is not intended for home use WARNING This devi...

Page 4: ...sting The recommended testing interval is once per year The protection categories against electric shock of the patient connections are FHR1 FHR2 TOCO MARK This symbol indicates that the instrument is...

Page 5: ...t a power cord with wet hands Make certain that your hands are clean and dry before touching a power cord WARNING The monitor should be installed by an authorized and qualified service engineer WARNIN...

Page 6: ...sing the equipment immediately after moving it from a cold environment to a warm humid location CAUTION Do not immerse transducers in liquid When using solutions use sterile wipes to avoid pouring flu...

Page 7: ...ring of two heart rates for twins This is achieved by using the facilities of two ultrasound transducers and an external contractions TOCO transducer with a recorder The monitor can display FHR TOCO M...

Page 8: ...al broad beam ultrasound transducer is used for monitoring fetal heart rate FHR1 The ultrasound transducer operates at a frequency of 1 0MHz Put the ultrasound transducer on maternal abdomen to transm...

Page 9: ...t Socket for FHR1 TOCO transducer and remote Marker FHR2 socket Socket for FHR2 Transducer Socket for Grounding Cable NET Socket for network Warning Symbol Push LCD Screen rotation lock 3 1 4 Main Int...

Page 10: ...nnel it is I under single fetal monitoring mode which is default I II selectable under twins monitoring mode it can be changed through the main menu 3 Volume level ranging from 0 7 0 stands for sound...

Page 11: ...setting parameters it contains alarm on off status alarm upper limit lower limit alarm postpone time print speed and print time 4 Indicate Bar In this item it includes system time bed No gestational a...

Page 12: ...m when FHR is in alarm range When symbol appears the alarm indicator status is shut off Press the button to enable audio alarm the alarm indicator becomes when FHR is in alarm situation the alarm soun...

Page 13: ...Function Freezing the screen Press the button to stop drawing and the screen becomes in frozen status press the button again to continue drawing This operation will clear the screen 8 Knob Key 1 Press...

Page 14: ...orm Review 5 PRINT SET Print Setup Enter Print Setup 6 MONI TYPE Monitor Mode Optional single twins the default is single fetus 7 LANGUAGE Language Selection Optional Chinese CH English EN the default...

Page 15: ...onds Note 1 When FHR is in alarm situation alarm indicator becomes red 2 When FHR exceeds the alarm limit and time exceed the set alarm delay time continuously alarm will occur and an alarm symbol wil...

Page 16: ...ed patient information 14 SAVE Save Information Save related patient information return to previous menu 15 Exit Return to previous menu Note When adjustable range is letter or number numeral and lett...

Page 17: ...r password under the item USR KEY before click CONFIRM This password is provided by manufacturer or distributor when Contec add new function to upgrade the system Version Revolving the knob key to ent...

Page 18: ...r works properly please read this chapter and Chapter 1 Safety Guidance And follow the steps before using the monitor 4 1 Open the Package and Check Open the package and take out the monitor and acces...

Page 19: ...r cabinet to check for a paper jam then feed the paper again Note Only use the manufacturer approved paper to avoid poor printing quality deflection or paper jam 4 5 Power on the Monitor WARNING If an...

Page 20: ...nsducer to the belt Apply aquasonic coupling gel to the face of the transducer 2 Acquiring the Fetal Heart Signal Determine the location of the fetal heart using palpation or a fetoscope Place the ult...

Page 21: ...the optimum position Positions with strong placental sounds swishing or fetal cord pulse indistinct pulse at fetal rate should be avoided If the fetus is in the cephalic position and the mother is sup...

Page 22: ...al movement At the moment the mark will show in the correspond position of trend wave The count of fetal movement will add 1 after each push of the button Push the button and hold for one second then...

Page 23: ...printing 5 3 Operation After Monitoring Remove transducers from patient Wipe transducer with a soft cloth to remove remaining ultrasound coupling gel Tear the paper at the folding place Switch off the...

Page 24: ...rface of the monitor Use dry and soft cloth to wipe WARNING Unplug the monitor from the AC power source and detach all accessories before cleaning Do not immerse the unit in water or allow liquids to...

Page 25: ...ng solution follow the manufacturer s directions carefully to avoid damaging the transducer 3 Wipe the transducer with a cloth soaked in water 4 Wipe the remained humidity with clean and dried cloth 6...

Page 26: ...tor Note The manufacturer has no responsibility for the effectiveness of controlling infectious disease using these chemical agents Please contact infectious disease experts in your hospital for detai...

Page 27: ...afety in Presence of Flammable Gases not suitable for use in presence of flammable gases Electromagnetic Compatibility Group Class A Mode continuous work Power Supply Working Voltage AC 100V 240V Freq...

Page 28: ...on the operation of pregnant women there will be a mark display in the bottom area of FHR wave display section FHR Alarm Alarm for high and low FHR which exceeds appointed limit A 1 2 Transducers 1 Ul...

Page 29: ...of FHR Symptom Possible Cause Solution Inconstant trace Inconstant display Wrong FHR No The pregnant woman is too fat No Improper ultrasound transducer position Change the position of ultrasound tran...

Page 30: ...rately and neither too tight nor too loose Maternal Movement Relax patient s spirit Fetal Movement Wait for a moment then monitor Too high TOCO sensitivity higher than 100 unit The body pressure from...

Page 31: ...Economic Technical Development Zone Hebei Province PRC Tel 86 335 8015433 Fax 86 335 8015432 E mail cms contecmed com cn Website http www contecmed com cn EC REPRESENTATIVE Shanghai International Hold...

Reviews: